Product Details

ESD Simulator Guns

Brand Name LISUN力汕
Place of Origin China
Model Number ESD61000-2

Product Features

What is a ESD test?
The static electricity generated by the human body to the object or between two objects may cause the electrical and electronic equipment circuits to malfunction or even be damaged. The ESD generator is designed for the withstanding ESD performance measurement for the assessment of electrical and electronic equipment. ESD61000-2/ESD61000-2A has a LCD touch screen in both English and Chinese. 

What is an esd simulator gun used for?
Electrostatic discharge simulator is highest electrostatic voltage can be up to 30 kv, which is enough to cover the most severe standard grade of electrostatic voltage requirement (the voltage requirements of the grade 4 air discharge electrostatic is 15 KV). The ESD test gun can be used for the most of electrical and electronic equipment for electrostatic discharge test, and also can ensure comparability and reproducibility of test.

Standard:
IEC 61000-4-2/EN61000-4-2/GB/T17626.2 “Electromagnetic compatibility (EMC) – Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test”
ISO10605 “Road Vehicles Test Methods for Electrical Disturbances from Electrostatic Discharge”
GB/T17626.2 
GB/T17215.322/IEC 62053-23:2003 “Electricity metering equipment (a.c.) – Particular requirements – Part 23: Static meters for reactive energy (classes 2 and 3)”

Specifications:

Product ModelESD61000-2ESD61000-2A
Output Voltage0.1~20kV±5%0.1~30kV±5%
PolarityPositive/NegativePositive/Negative
Energy Storage Capacitance150pF±10% (Replaceable)150pF±10% (Replaceable)
Discharge Resistor330Ω±5% (Replaceable)330Ω±5% (Replaceable)
Current Rise Time0.6~1ns0.6~1ns
Testing FunctionsSingle, Count, 20pps, Air, Contact, IEC levelSingle, Count, 20pps, Air, Contact, IEC level
Trigger ModeMANUAL/AUTOMANUAL/AUTO
Number of Discharge1~99991~9999
Repetition0.05s~99.99s0.05s~99.99s
Working PowerAC220V (Option 110V) ±10%, 50/60HzAC220V (Option 110V) ±10%, 50/60Hz


Calibrate Report:

Output Voltage (kV)First Peak Current (A)Current at 30ns (A)Current at 60ns (A)Rise Time (ns)
27.294.102.200.93
415.407.904.300.97
623.2012.106.500.97
829.4016.209.300.89
-27.393.502.300.92
-415.507.704.300.89
-623.4011.906.300.90
-831.8016.108.200.90