Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar
Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar
Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar
Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar
Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar

Saudi Arabia filling liquid nitrogen tank KGSQ pressurized liquid nitrogen dewar

Negotiable /Set

Min.Order:1 Set

Supply Ability:
500 Set / Sets per Month
Port:
Qingdao
Payment Terms:
T/T L/C D/P D/A PayPal
Delivery Detail:
7 days

Quick Details View All >

Brand Name:
KGSQ
Place of Origin:
China
Model Number:
YDZ-200
Condition:
New

Henan Tianzhidao Biological Technology Co., Ltd.

Business license
Business Type: Trading Company
Main Products:

Product Details

The stainless steel liquid nitrogen tank is also a self-pressurizing liquid nitrogen tank. The tank itself has a pressurizing system and is adjusted through a pressurizing valve. When the booster valve opens, the pressure increases, and when it closes, the pressure decreases. During use, what is the reason why the booster valve is opened but unable to boost pressure?

The reasons why the pressure does not rise after the booster valve of the self-pressurized liquid nitrogen tank is opened are as follows:

1. The booster valve is damaged

When the vent valve and the inlet and outlet valve are closed, and the pipeline is unobstructed but does not pressurize, it means that there is a problem with the booster valve itself and the booster valve needs to be replaced. You can contact the manufacturer for replacement.

2. Pipe blockage

The main reason for pipeline blockage is that external air and moisture enter and are not discharged in time. Under the influence of low temperature in the tank, they condense into ice and block the pipeline.

The correct way to clear the pipe is to use hot nitrogen or hot air to blow away the ice and clear the pipe. After unblocking, it is recommended to transport liquid nitrogen to test whether it is unblocked. Do not use tapping to unblock.

3. Insufficient pressure

Check whether the pressure in the liquid nitrogen tank is sufficient. If the pressure in the liquid nitrogen tank is too low, the pressure cannot be increased even if the pressure increase valve is opened.

4. Container leakage

The container leaks and the vacuum is destroyed, which will also result in the inability to pressurize.

It is easy to judge whether the container is leaking. You can observe the frosting phenomenon on the tank body. The location of a large area of frost is the leakage location.

At this time, the jar can no longer be used normally. You can contact the manufacturer for after-sales treatment.

In short, there are many reasons why the pressure does not rise after the booster valve of the self-pressurizing liquid nitrogen tank is opened. Check them one by one to find out and solve the problems.

If you have any needs, please contact KGSQ liquid nitrogen tank:

Email: tcln2container@126.com

WhatsApp: +86 18903712477

Official website:

/YDZ-Series-Liquid-Nitrogen-Supply-Tank.html


Contact Supplier

Jian Yang Chat Now
Telephone
86-0371-56788280
Mobile
86-18903712477
Address
High-tech Industrial Development Zone of Zhengzhou, 289 West Third Ring Road, Building 11, 1305 Ji'nan,Shandong

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Machinery > Other Machinery & Industry Equipment
Categories
Hi~