300W photocatalytic Xenon lamp light source water splitting
USD $4500 - $5000 /Set
Min.Order:1 Set
Beijing China Education Au-light Co., Ltd.
Xenon Light source is widely used for scientific research in fields of Photolysis of Hydrogen and Oxygen, CO2 Reduction, Light and Heat Synergy, Photothermal Catalysis, Photochemical Catalysis, Photochemical synthesis, Photodegradable pollutants, Water Pollution Treatment, Biological Lighting, Optical Inspection, Solar Simulation.
Advantages
CEL-PF300-T6 adopt to new outlooking, modularity heat dissipation system, improve the stability of light output
Adopt to new developping and upgraded power supply and high voltage trigger module.
After turn off, the main fan heat dissipation keep 60 s delay to protect the lifetime well.
PE300 lamp can be replaced quickly and easily with no extra complicate assembly work.
New light path structure, combined lamp, heat disspation, Isolation, flange multi concentric.
New temperature control feedback system, high voltage isolation,improve the accuracy.
Support all M62 M52 optical filters. Multiple filters can be mounted.
Support optional automatic lifting platform LMP400 to adjust the height easily.
Support optional photocatalysis experiment cabinet.(400*400*700mm, can load 2 sets of xenon light source)
Technical Parameters
CEL-PF300-T6 | CEL-PF300UV-T6 | |
Output light Power | 0~20 Sun(1Sun=1000W/m2) | 0~20 Sun |
Spectral Output(nm)
| 300nm~2500nm(ozone free) | 200nm~2500nm(ozone) |
Beam Diameter | >=60mm | >=60mm
|
UV Output,<390nm (Watts) | 2.6W | 6.6W |
IR Output, >770nm (Watts) | 28.8W | 26.8W |
Visible Output, 390-770nm (Lumens) | 5000Lu、18.6W、5600K | 4500Lu、16.6W、5050K |
Input Power(Watts) | 300W(Igniter DC 30KV,Working DC 14V) | |
Working Current (Amps DC) | 21A(10A~22A) | |
Total Output light Power | 50W | |
Bulb Life(Hours) | >1000H,Ultimate 6000H | |
Output stability | ±1% | |
Temperature Control System | The light source system adopts multi-point temperature monitoring to ensure the stable output of the light source; The fan speed delay is automatically adjusted according to the system temperature to stabilize the light intensity output; Fan delay 60 s turn off | |
Options | CEL-LMP400 Lifting Platform CEL-LB70 Photochemical experiment box CEL-NP2000-2 Glare Optical Power Meter | |
Optional quartz coated filter(M62) | General Optical Filters AREF(Full Reflective,200-2500nm) - Default included with lamp set VisREF (Reflect Visible Light, 350-780nm) UVREF (Reflect UV, 200-400nm) AM1.5G (300-1100nm) UVIRCut400 (Band Pass 400-780nm) UVIRCut420 (Band Pass 420-780nm) CUT800 (Pass Above 800nm) Band Pass Filters QD254,QD275, QD280, QD295,QD300-800QD313, QD320, QD325, QD330, QD334, QD350, QD355, QD360, QD365, QD370, QD375, QD380,QD400,QD405, QD420, QD435, QD450,QD475,QD500,QD520, QD550, QD578,QD600,QD630, QD650, QD670,QD700,QD730, QD765,QD850,QD940 |