Dust Proof Environmental Test Chamber Anti Explosive
Negotiable /Bag
Min.Order:1 Bag
Guangdong ALI Testing Equipment Co,.Ltd
Advantage of Environmental Test Chamber:
● Interior of Chamber is SUS304 stainless steel.
●The air velocity for the blowing of dust is controlled by a dust blowing mechanism that includes a motor and blower combination
●Dust re-cycle blowing by circulating pump.
●Dust collection section is located on bottom of test chamber.
Air conditioning system:
1, the air control: forced ventilation circulation, balance thermostat method (BTC). This method not only refers to the case of the cooling system for continuous operation, the control system according to the temperature set point of the heater to control the output of the results by automatically calculating PID output, and ultimately achieve a dynamic equilibrium.If you need more models about our test equipment, please reference our chamber-testing(.)com
2, air circulating means: built between the air circulation fan duct and the long axis of stainless steel, the use of energy efficient refrigerators and regulation system, by effective and efficient heat exchange ventilator, to achieve the purpose of the temperature change. By adjustable air outlet adjusting bolt, thereby significantly improving the chamber temperature uniformity;
3, the air cooling method: multi-stage diaphragm air heat exchanger.
4, air heating: Watkins quality nickel chrome electric heater.
Features:
1, The inner chamber is made of mirror stainless steel.
2, Semicircular corners inner chamber is easy to clean.
3, Shelf can be adjusted.
4, LED microcomputer controller make the temperature and humidity constant.
5, Touch key parameter setting.
6, Timer, over temperature alarm function.
7, Large angle observation window, easy to watch the goods in the inner chamber with lighting.
8, Large-capacity humidifier, easy to use
Performance parameters:
Internal dimensions (mm) | 800*1000*1000 | 1000*1000*1000 | 1000*1500*1000 | 1000*2000*1000 |
Overall dimensions (mm) | 1040*1450*1960 | 1200*1450*1960 | 1250*1900*1960 | 1300*2400*1960 |
Temperature range | Ambient to +50℃ | |||
Relative Humidity | <30%RH | |||
Normal wire diameter | 50um | |||
Normal width of a gap between wire | 75um | |||
The Talcum powder amount | 2kg ~ 4kg/m3 | |||
Fight time | 0 ~ 99H59M | |||
Fan cycle time | 0 ~ 99H59M | |||
Specimen power outlet | Dust-proof socket AC220V 16A | |||
Controller | programmable color display touch screen controller | |||
PC Link, R-232 interface | ||||
Vacuum system | Equipped with a pressure gauge, air filter, pressure regulator, connecting tube | |||
Exterior material | Steel Plate with protective coating | |||
Interior material | SUS304 stainless steel | |||
Safety System | Electric leakage, short circuit, Over-temperature, motor overheating Over-current protection/ Power failure memory function for controller | |||
Power supply | 380V 50HZ | |||
Standard | IEC60529, ISO20653 |
Customer Support Services:
Ø Installation
Ø Training (Training customer employees)
Ø Calibration
Ø Preventative maintenance
Ø Replacement parts
Ø Assistance via phone or internet