Filter integrity testerV4.0
Negotiable /Set
Min.Order:1 Set
Beijing Neuronbc Laboratories Co.,Ltd.
Features
u Automatic air inlet design, to ensure efficiency and test accuracy at the same time;
u The instrument actually measures the up stream volume, so the test result is more accurate;
u Domestic first model to introduce water immersion method;
u Widely used by customers, with a high market share and cost performance;
u Automatic control of air intake level, with a smooth and safe air intake, to ensure the service life of filter element;
Parameters:
Power Supply | 100 – 260 VAC, 50HZ, 120W ; |
Operation pressure | 100-8000mbar (115psi) ; |
Unit | mbar |
Operation Condition | Ambient Temperature:+5℃ ~ +40℃;Relative Humidity:10-80% |
Dimension(mm) | 400(Depth)*240(Width)*270(Height1)*220(Height2) |
Test Function | Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test;Diffusion Flow Test;Water Immersion Test; |
Test Accuracy | Upstream Volume:± 4%;Bubble Point:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion Test:≤± 0.05ml |
Test Scope | Bubble Point :590-6900mbar Diffusion Flow Speed:1-180ml/min WI:0.01-30ml/min |
Applied Range | Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (3 core within 20 inches) |
Print Function | Built-in printer; |
Historical Records | 500 sets; |
Display Screen | 5.7″TFT,Single color; |
Serial Connection | RS232; |
Language | Chinese/English; |
Work Mode | Online/Offline; |
Applicable Environment | Above B level; |
Weight | 7.5KG; |