BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette
BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette
BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette
BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette
BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

Negotiable

Min.Order:1

Supply Ability:
1000 per Month
Payment Terms:
T/T Credit Card

Hangzhou Boyn Instrument Co., Ltd.

Business Type: Agent
Main Products: , , ,

Product Details

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette       

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

 

Features(Single-channel)
Light weight, ergonomic, traditional design
Digital display clearly reads volume setting
Pipettes cover a volume range from 0.1μL to 10mL
Easy calibration and maintenance
Each TopPette supplied with an individual calibration certifificate according to ISO8655
The lower section can be autoclaved
 
Features(Multi-channel)
8 and 12 channel pipettes are available for standard 96-well plate
Dispensing head rotates for optimum pipetting convenience
Individual piston and tip cone assemblies allow easy repair and maintenance
Compound material-made tip cone secures high sealing performance
Can be used with universal style pipette tips
 
 

Specification

 

Single-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume Error limits in accordance with ISO8655-2
(Accuracy error) (Precision error)
%  μL % μL
0.1-2.5μL 0.05μL 2.5μL 2.50% 0.0625 2.00% 0.05
1.25μL 3.00% 0.0375 3.00% 0.0375
0.25μL 12.00% 0.03 6.00% 0.015
0.5-10μL 0.1μL 10μL 1.00% 0.1 0.80% 0.08
5μL 1.50% 0.075 1.50% 0.075
1μL 2.50% 0.025 1.50% 0.015
2-20μL 0.5μL 20μL 0.90% 0.18 0.40% 0.08
10μL 1.20% 0.12 1.00% 0.1
2μL 3.00% 0.06 2.00% 0.04
5-50μL 0.5μL 50μL 0.60% 0.3 0.30% 0.15
25μL 0.90% 0.225 0.60% 0.15
5μL 2.00% 0.1 2.00% 0.1
10-100μL 1μL 100μL 0.80% 0.8 0.15% 0.15
50μL 1.00% 0.5 0.40% 0.2
10μL 3.00% 0.3 1.50% 0.15
20-200μL 1μL 200μL 0.60% 1.2 0.15% 0.3
100μL 0.80% 0.8 0.30% 0.3
20μL 3.00% 0.6 1.00% 0.2
50-200μL 1μL 200μL 0.60% 1.2 0.15% 0.3
100μL 0.80% 0.8 0.30% 0.3
50μL 1.00% 0.5 0.40% 0.2
100-1000μL 5μL 1000μL 0.60% 6 0.20% 2
500μL 0.70% 3.5 0.25% 1.25
100μL 2.00% 2 0.70% 0.7
200-1000μL 5μL 1000μL 0.60% 6 0.20% 2
500μL 0.70% 3.5 0.25% 1.25
200μL 0.90% 1.8 0.30% 0.6
1000-5000μL 50μL 5000μL 0.50% 25 0.15% 7.5
2500μL 0.60% 15 0.30% 7.5
1000μL 0.70% 7 0.30% 3
2-10mL 0.1mL 10mL 0.60% 60 0.20% 20
5mL 1.20% 60 0.30% 15
2mL 3.00% 60 0.60% 1

 

Fixed Volume Pipettes
Volume Range Increment Test Volume Error limits in accordance with ISO8655-2
(Accuracy error)  (Precision error)
% μL  % μL
5μL  - 5μL 1.30% 0.065 1.20% 0.06
10μL  - 10μL 0.80% 0.08 0.80% 0.08
20μL  - 20μL 0.60% 0.12 0.50% 0.1
25μL  - 25μL 0.50% 0.125 0.30% 0.075
50μL  - 50μL 0.50% 0.25 0.30% 0.15
100μL  - 100μL 0.50% 0.5 0.30% 0.3
200μL  - 200μL 0.40% 0.8 0.20% 0.4
250μL  -  250μL 0.40% 1 0.20% 0.5
500μL  - 500μL 0.30% 1.5 0.20% 1
1000μL  - 1000μL 0.30% 3 0.20% 2
2000μL  - 2000μL 0.30% 6 0.15% 3
5000μL  - 5000μL 0.30% 15 0.15% 7.5

 

12-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume Error limits in accordance with ISO8655-2
(Accuracy error)  (Precision error)
% μL % μL
0.5-10μL 0.1μL 10μL 1.50% 0.15 1.50% 0.15
5μL 2.50% 0.125 2.50% 0.125
1μL 4.00% 0.04 4.00% 0.04
5-50μL 0.5μL 50μL 1.00% 0.5 0.50% 0.25
25μL 1.50% 0.375 1.00% 0.25
5μL 3.00% 0.15 2.00% 0.1
50-300μL 5μL 300μL 0.70% 2.1 0.25% 0.75
150μL 1.00% 1.5 0.50% 0.75
50μL 1.50% 0.75 0.80% 0.4

 

8-channel Adjustable Volume Pipettes
Volume Range Increment Test Volume Error limits in accordance with ISO8655-2
(Accuracy error)  (Precision error)
% μL % μL
0.5-10μL 0.1μL 10μL 1.50% 0.15 1.50% 0.15
5μL 2.50% 0.125 2.50% 0.125
1μL 4.00% 0.04 4.00% 0.04
5-50μL 0.5μL 50μL 1.00% 0.5 0.50% 0.25
25μL 1.50% 0.375 1.00% 0.25
5μL 3.00% 0.15 2.00% 0.1
50-300μL 5μL 300μL 0.70% 2.1 0.25% 0.75
150μL 1.00% 1.5 0.50% 0.75
50μL 1.50% 0.75 0.80% 0.4

 

                                                                                                                                                                                                                        Please send inquiry>>

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

                                       If you need any detail packing information, please contact us>>

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

                                                                       If you have any question, please contact us>>

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

BN-TopPette non-autoclavable variable Single-channel/Fixed volume/8-channel/12-channel micro pipette micropipette

Contact Supplier

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Chemicals > Lab Supplies > Pipette
Categories
Hi~