SX736 SX736Portable Ph/Conductivity/DO pH high purity water and pure water with ammonia.Meter
Negotiable
Min.Order:1
SX725 Portable pH/Dissolved Oxygen Meter With Data Storage USB Data Output
Laboratory equipment DGB402F/403F Portable Intelligent Chlorine Meter with Reliable Quality
Cheap direct reading water biochemical oxygen demand analyzer bod meter
USB Intelligent Mercury Free Biochemical Oxygen Demand Meter BOD Analyzer
Introduction
Multi-parameter meter, equipped with pH electrode, conductivity electrode and dissolved oxygen electrode, automatic identification of electrode by meter. |
Built-in microprocessor chips, intelligent functions such as automatic calibration, automatic temperature compensation, data storage, function set-up, self-diagnose information, automatic power-off and low voltage display. |
Automatic identification of 13 pH standard buffer solutions, three options of standard buffer solutions: Europe and America series, NIST series and China series. |
Able to set pH measurement modes of high purity water and pure water with ammonia. |
Automatic identification of 8 conductivity standard calibration solutions, two options of standard calibration solutions: Europe and America series and China series. |
Adopt advanced conductivity measurement technology, use conductivity electrode of constant K=1, only need one point calibration and meet the measurement precision requirement of 0.5μS/cm ~ 200 mS/cm. |
Ability to switch among conductivity, TDS, salinity and resistivity. Multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale. |
Conductivity measurement mode has the function of non-linear temperature compensation for high purity water. |
Measurement mode of Dissolved oxygen has the function of automatic temperature compensation, automatic salinity compensation and manual air pressure compensation. |
New polarographic type dissolved oxygen electrode can measure temperature, salinity and dissolved oxygen simultaneously, equipped with special dissolved oxygen electode calibration set, polarization only for 3~5 min. |
Meter meets IP57 waterproof rating, equipped with calibration solution and portable case. |
Specifications
Category | Product Name | Basic features and specifications |
PH |
Measuring range |
(-2.00 ~ 19.99)pH |
Resolution |
0.1/0.01 pH | |
Accuracy |
Meter:±0.01pH; Overall: ±0.02pH | |
Input current |
≤2×10-12 A | |
Input impendance |
≥1×1012 Ω | |
Stability |
±0.01 pH/3h | |
Temp. compensation range | (0 ~ 100)°C(automatic) | |
ORP |
Measuring range (mV/ORP/EH) |
-1999mV ~ 0 ~ 1999mV |
Resolution |
1mV | |
Accuracy |
Meter:±0.1% FS, Overall: ±15mV | |
Conductivity | Measuring range | Conductivity:(0.00~19.99)μS/cm (20.0~199.9)μS/cm (200~1999)μS/cm (2.00~19.99)mS/cm (20.0~199.9)mS/cm TDS: (0 ~ 100)g/L Salinity: (0 ~ 100)ppt Resistivity:(0 ~ 100)MΩ·cm |
Resolution |
0.01/0.1/1μS/cm 0.01/0.1 mS/cm | |
Accuracy | Meter:±1.0% FS, Overall:±1.5% FS | |
Temp. compensation range |
(0 ~ 50)°C(manual) | |
Electrode constant |
0.1 / 1 / 10 cm-1 | |
Benchmark temperature |
25°C,20°C and 18°C | |
Dissolved Oxygen |
Measuring range |
(0 ~ 40.00)mg/L(ppm) (0 ~ 200.0)% |
Resolution |
0.1/0.01 mg/L(ppm) 1/0.1 % | |
Accuracy |
Meter:±0.10 mg/L, Overall:±0.40 mg/L | |
Response time |
≤30 s(25°C, 90%response) | |
Residual current | ≤0.1 mg/L | |
Temp. compensation range | (0 ~ 45)°C(automatic) | |
Salinity compensation range | (0 ~ 45)ppt(automatic) | |
Atmospheric pressure compensation range |
(66 ~ 200)kPa (manual) | |
Automatic calibration |
Air saturated by water; water saturated by air | |
Electrode type |
Polarogram type | |
Other Technical Parameters |
Data storage |
200 groups |
Storage content | Series number, measuring value measuring unit and temperature, | |
Power |
Two AA batteries (1.5V x2) | |
Size and weight |
meter:(65×120×31)mm/180g;Suitcase: (360×270×76)mm/1.6Kg | |
Quality and safety certification | ISO9001:2000, CE and CMC | |
Working Condition |
Environment temperature |
5 ~ 35 °C(0.01grade) |
Environmental humidity |
≤85% | |
IP rating |
IP57 Dustproof and waterproof |
Standard Kit Includes
1. Model SX736 pH/mV/Conductivity/DO meter | 1set |
2. 201T-S plastic pH/ATC three-in-one combined electrode | 1pcs |
3. 2301-S plastic conductivity electrode | 1pcs |
4. DO500 DO electrode | 1pc |
5. pH standard buffer solution(pH4.00, pH7.00 and pH10.01/50mL) | 1btl/each |
6. 1413μS/cm conductivity standard solution (50mL) | 1bottle |
7. DO503 membrane cap (DO electrode use)) | 3pc |
8. DO502 DO electrode inner solution (30mL) | 1bottle |
9. Cathode polishing paper | 3pcs |
10. Screw driver | 1pcs |
11. Spare AA batteries | 2pcs |
12. Operation manual | 1pcs |
13. Carrying case | 1unit |
Certificate