Glass pasteur pipette with single channel
Electronic Pipette:
The pipette is available in single-channel models covering volume range of 0.5 to 5000μl and muti-channel models from 0.5 to 1200μl.
Features:
- Comprehensive range of liquid handling protocols with easy programming
- DC motor with build-in error control improves pipetting precision and provides more reliable results
- Efficient lithium-ion battery offers long runtime on each charge
- 9 speeds for aspiration and dispensing
- Auto-clavicle low part
- Wireless charging
Specifications:
Channels |
Volume Range |
Increment |
Test Volume |
Inaccuracy |
Imprecision |
||
μl |
μl |
% |
s.d.*μl |
CV%* |
|||
1 |
0.5-10μl |
0.01μl |
10 |
±0.10 |
±1.0 |
0.05 |
0.5 |
1 |
5-50μl |
0.1μl |
50 |
±0.40 |
±0.8 |
0.15 |
0.3 |
1 |
30-300μl |
1μl |
300 |
±1.8 |
±0.6 |
0.6 |
0.2 |
1 |
100-1000μl |
1μl |
1000 |
±6.0 |
±0.6 |
2.0 |
0.2 |
1 |
1000-5000μl |
- |
5000 |
±30.0 |
±0.6 |
10.0 |
0.2 |
8 |
0.5-10μl |
0.1μl |
10 |
±0.24 |
±2.4 |
0.16 |
1.6 |
8 |
5-50μl |
0.1μl |
50 |
±0.75 |
±1.5 |
0.35 |
0.7 |
8 |
30-300μl |
1μl |
300 |
±3.0 |
±1.0 |
0.9<br |
Mechanical Pipettes( Adjustable and fixed volume)
The quality control involves gravimetric testing of each pipette with distilled water at 22 degree.
Top-pipette Single-channel Adjustable Volume Pipettes |
|||||||
Cat No. |
Volume Range |
Increment |
Test Volume |
Maximum permissible systematic error (Inaccuracy) |
Maximum permissible random error (Imprecision) |
||
% |
μl |
% |
μl |
||||
711111010000 |
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
|||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
|||
711111040000 |
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
|||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
Certificate