S7160CIO2-H  Online Chlorine Dioxide Monitor
S7160CIO2-H  Online Chlorine Dioxide Monitor
S7160CIO2-H  Online Chlorine Dioxide Monitor

S7160CIO2-H Online Chlorine Dioxide Monitor

280 - 380 /Piece

Min.Order:1 Piece

Supply Ability:
20000 Piece / Pieces per Month
Port:
Shanghai
Payment Terms:
T/T L/C Cash

Quick Details View All >

Brand Name:
lanben/OEM
Place of Origin:
China
Model Number:
S7160CIO2-H

Suzhou SKYING Control Technology Co., Ltd.

Business license
Main Products: Digital sensor electrode , digital controller ,Water quality multi-parameter monitor ,pH/ORP meter , conductivity , resistivity , TDS , salinity meter ,Dissolved oxygen meter , fluorescent dissolved oxygen meter ,Mud level gauge , hardness monitor ,pH meter , residual chlorine meter , sulfur dioxide monitor ,Suspended solids (sludge) concentration measuring instrument ,Portable dissolved oxygen meter , pen ph meter ,Digital pH Digital Conductivity Digital Dissolved Oxygen ,Benchtop pH Benchtop Dissolved Oxygen Benchtop Conductivity ,Portable pH Portable Dissolved Oxygen Portable Conductivity ,Pen ph Pen Dissolved Oxygen Pen Conductivity

Product Details

7160CIO2-H  Online Chlorine Dioxide Monitor


Applications

Industrial online chlorine dioxide monitor is a water

quality online monitor with microprocessor.

It is widely used in drinking water treatment plants,

drinking water distribution networks, swimming pools,

water treatment projects, sewage treatment, water

disinfection and other industrial processes.

It continuous monitoring and control chlorine dioxide

& temperature value in water solution.


Feature

● LCD Display.

●Intelligent menu operation.

●Data Recording & Curve Display.

●Manual or automatic temperature compensation.

●Two sets of relay control switches.

●High & low alarm, and hysteresis control.

●4-20mA&RS485Multiple output modes.

●Display measurements, temperature, state, etc.on the same interface.

●Password protection function to prevent misoperationby non-staff.


Technical parameters:

Measurement range:Chlorine dioxide:0~20.00mg/L;Temperature:0~50.0ºC;

Resolution:Chlorine dioxide:0.001mg/L;

Temperature:0.1ºC;

Basic error:Chlorine dioxide:±1%F.S;

Temperature:±0.3ºC;

Response time:90%less than 90s;

Current output:0/4~20mA(load resistance<750Ω);20~4mA(load resistance<750Ω);

Communication output:RS485 MODBUS RTU;

Two sets of relay control contacts:3A 250VAC,3A 30VDC;

Power supply(selection):85~265VAC±10%,50±1Hz,power consumption≤3W;9~36VDC,power consumption≤3W;

Dimensions:98x98x130mm;

Installation: Panel & wall mounting;

Panel opening size:93x93mm;

Protection Level:IP65;

Weight:0.6kg;

Working environment:Operating temperature:-10~60ºC;

Relative humidity: ≤90%;

No strong magnetic field around exceptthe geomagnetic field.



Contact Supplier

Mr. andi@holigroup.cn andi@holigroup.cn Chat Now
Mobile
86-13151156165
Skype
chaochina@126.com
Whatsapp
86 13151156165

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Measurement & Analysis Instruments > Analyzers > PH Meters
Categories
Hi~