LED display Display Type and 220V Power Supply load cell indicator 2 relay output



Specifications: 

1. Analog input

Analog inputs use 20 bits AD chip, Count Speed up to 80 times / sec, the measurement accuracy of 0.2%

Display range -9999 ~ 19999

The largest sub-degree 50000

For the bridge voltage 5V, 100mA

2. The switch output

3 relay or OC outputs. Relay contacts AC 250V / 1A, OC output drive 100mA / 48V optional.

3. Transmission output

Output 4-20mA, 0-10V, 12bit precision driving 500 ohm load. Optional features.

4. communication port

485,232 communication interface, upload executable modbusRTU and active protocol. This function is optional. Only can choose one function with transmission function

5. A simple comparison output

Features: 3 output meter can be set to be greater than or less than the comparative comparison. When the measured value exceeds the upper limit output is active when the measured value is less than the upper limit minus hysteresis output is invalid. If the output is less than comparison, the measured value is less than the lower limit of the output is active when the measured value is greater than the lower limit plus hysteresis invalid.

Input signal
0-2.5MV/V Excitation Voltage 5V/100mA
Measurement accuracy
0.1%
Sampling rate
10 times or 80 times/second can be set
Display mode
Single row red LED display. Main display window display range: -9999~29500
On-off input
1 way input, short circuit is effective.(selective assembly)
Transmitting
Output 4-20mA, 0-20mA, 0-5V, 0-10V, 12bit accuracy, driving load≤500Ω.
Communication
485com port can execute Modbus RTU protocol or Active upload protocol
Working Environment
environment temperature: -20~50℃; relative humidity≤85%RH; Avoid strong caustic gas.
Working power supply
AC 100~240V, 50~60HZ; power consumption≤5W
Structure
Panel mounting / cabinets mounting, high quality plastic shell, dimension 96*48*84mm,Cut size:92×45mm

Operation:

Menu Structure:


3.1 parameter modification. Standby mode, press [k1] + password to enter the password written 100. Entering advanced parameters (13 parameters later) password is written 105.

Specific steps are as follows: Press [k1] key, there will be 00000, after press [k4] key to move the blinking digit moves to the flashing digit to be modified, press [k3] to modify the blinking digits, password changes is correct, press [k1] key to enter the set parameters, and then press [k1] to switch the set of parameters, select the parameter to be modified, press [k4] enter the parameters after the modification is completed, press [k1] OK to return to the measurement interface [k2]

3.2 weight calibration. Standby mode, press [k2], to see details of the third page (V: calibration method)

3.3 Key cleared. Standby mode, press [k4], and the measured value is less than zero range is completed is cleared.

3.4 restore the factory settings. 3.1 according to enter parameter modification (but need to enter a password to use special password) to switch to the first 42 parameters, enter a number between 11-19, and then press the [k1] to switch to the next parameter, press [k2 [to exit the way to restore the instrument parameters 1-9 initial state.



Display symbolically

A 1 Alarm output value

b 2 alarm output value

c 3 alarm output value

d alarm hysteresis

AL alarm status

st running state

E A / D code value

dA transmitter output code

U Software version


Calibration method
Standby mode, press [k2], appears E0xxxx, press [k4] is cleared, there E00000 zero calibration is complete. Press [k2] qualifying C00000, plus a known standard weights or heavy objects, press [k4], and then by pressing [[k3] to increase the value + [k4] to move the blinking digit, modified weight by weight, stable, press [k2] OK, the calibration is complete, the value and weight of the weights should be consistent, if not the exact same order again once calibrated, accurate after calibration, press [k1] return force measurement interface.

Related Product

CALT DYLF-102 Spoke type load cell

CALT DYLY-103 S type load cell

CALT DYHW-116 Miniature load cell

CALT DYX-306 Aluminium alloy Cantilever beam load cell

CALT DYX-301 Cantilever beam load cell

CALT DYMH-106 Miniature load cell

Detailed Image
certifications
About Us

COMPANY PROFILE

Shanghai QiYi Electrical & Mechanical Equipment Co., Ltd is a technology industrial enterprises integrating R & D, production, agency and sales. our company focus on the rotary encoders, displacement sensors and proximity sensor with CALT brand. We have many years of experience in supplying rotary encoders and sensors, which widely used in CNC machine, steel rolling machinery, linear guide systems, hydraulic systems, lifting machinery, turbine and other multiple industries. Located in Shanghai, we enjoy convenient air, water and land transportation.

FAQ

Contact Supplier

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Measurement & Analysis Instruments > Weighing Scales
Categories
Hi~