Product Details

Automatic Abel Closed Cup Flash Point Tester ISO 13736 Abel flash point apparatus

Brand Name ADDITION INSTRUMENT
Place of Origin China
Model Number JF21789
Power Electronic
Usage Other, flash point apparatus

Product Features

Technical Parameter

1. Applicable standards: ISO 13736, IP170, ISO 1516, ISO 1523, GB/T 21789

2. Instrument form: desktop (all stainless steel shell)

3. Test range: -30℃~80, temperature control accuracy ±0.1

4. Refrigeration source: built-in high-quality semiconductor refrigeration metal bath cold trap

5. Stirring mode: 030rpm (adjustable)

6. Repeatability: <±1

7. Temperature detection sensor: PT100 thermocouple sensor

8. Flash point detection sensor: CRC detection thermocouple

9. Display: Color touch LCD display.

10. Ignition method: automatic ignition (choose one of gas ignition and electronic ignition method), which is determined by high-sensitivity thermocouple detection method.

11. Display content:

Control temperature and expected flash point setting temperature.

Fault prompt screen.

Each model parameter and air pressure setting.

12. Power consumption: 800W

13. Working power supply: AC220,240V 50,60HZ

14. Dimensions: 400×380×350mm

15. Weight: 20kg

 

 Main Feature

1. The experimental mode is selected, and the flash point detection test is carried out in strict accordance with the operating procedures of different Abel closed cup (Abel closed cup) standards, and the flash point value of the sample is detected:

A. ISO13736, ISO1516, ISO1523, GB/T 21789 and other test modes selection.

B. Expected flash point search mode: perform a quick test on the sample for which the expected flash point value is unknown, and retrieve the expected flash point

C. User-defined.

(Different standards have different requirements for temperature rise control and ignition detection mode, users can set the test mode by themselves)

2. Set the flash point test parameters according to the operation mode on the LCD screen, press the START button, and the whole test process will be carried out automatically.

3. Color touch LCD screen, intuitive operation and display the whole process of flash point test. The interface is simple, directly click to set the experimental mode, easy to use and reliable, and the test process and measurement data are clear at a glance.

4. The instrument is compact, small and beautiful in appearance.

5. The instrument has self-fault diagnosis and alarm functions.

 

 


You May Like

Find Similar Products By Category

You May Like

View More
Chat Now Contact Now