Product Details

WML-420A VET Veterinary 3-diff Auto Hematology Analyzer

Brand Name VET Veterinary 3-diff Auto Hematology Analyzer
Place of Origin China
Model Number WML-420A

Product Features

WML-420A VET Veterinary 3-diff Auto Hematology Analyzer

Features 

Stable hematology analyzer combined with reliable hardware and intuitive software,for sure it is designed to be a easy operation diagnostic system with convenient data communicate interface,excellent performance.

Animal species: Canine, Bovine, Feline, Ovine, Equine, Simian, Porcine, Murine, and 4 user-defined species

Principle: Impedance for WBC, RBC, PLT counting Compact design 

Throughput: 35 samples per hour

3-part differentiation of WBC, 20 parameters +3 histograms

3 counting modes: whole blood, capillary blood and prediluted

Micro sample volume: 10 ul

Built-in thermal printer

Reagent:Diluent,Lyse,PB Cleanser,E-Z Cleanser

12months warranty

General Features 

Model 420A VET  

Throughput 35 samples/hour  

Channels 1  

Aperture diatmeter NO 

Principle : Impedance for WBC,RBC,PLT;SFT method for HGB20 parameters(including WBC, Neu#, Lym#, Mid#, Neu%,Lym%, Mid%,

ParametersRBC, HGB, HCT, MCV, MCH,MCHC,RDW-SD,RDW-CV, PLT, MPV, PDW, PCT, P-LCR)

Histograms3 histograms(including WBC,RBC,PLT histograms) 

HGB lampLED 540nm 

Sample volume Normal: 10ul, Prediluted: 20ul

Standard reagent 

Reagent Standard: Dilent: 20L, Lyse: 500ml 

Cleanser Standard: Probe cleanser: 50ml

Software 

Calibration   Manual & auto calibration 

Counting modes Venous, capilary, prediluted 

Print   Auto print, manual print 

LIS   Support LIS 

Language   English, Spanish,Italian,Portuguese, etc

Specifications 

Input 8 inch touch screen 

Output Internal printer, support external printer 

Printer paper 80×35mm 

Interface RS232 port, USB port, VGA port 

Storage 30,000 results with histogram 

Dimension 565mm×467mm×571mm

Blockage clear High voltage, high pressure flush 

Net weight 19 KG

Performance

   Carry-over rate

Parameter

CV

WBC

≤ 2%

RBC

≤ 1%

HGB

≤ 2%

PLT

≤ 2%

 

 

 

                             Background

Parameter

Background

WBC

≤ 0.2×109/L

RBC

≤ 0.02×1012/L

HGB

≤ 1 g/L

HCT

≤ 0.5%

PLT

≤ 10×109/L

 

 

 

                         

                         Precision

Parameter

CV

Measurement range

WBC

≤ 2.0%

(4.0-15.0×109)/L

RBC

≤ 1.9%

(3.5-6.0×1012)/L

HGB

≤ 1.9%

(110.0-180.0g)/L

MCV

≤ 0.4%

(80.0-110.0)fL

PLT

≤ 4.0%

(100.0-500.0×109)/L

 

 

 

 

                          Linearity

Parameter

Measurement range

CV

WBC

(0-5.9×109)/L

(±0.3×109)/L

6.0-99.9×109)/L

±5%

RBC

(0-0.99×1012)/L

(±0.05×1012)/L

(1.0-9.99×1012)/L

±5%

HGB

(0-99.0g)/L

(±2.0g)/L

(100.0-300.0g)/L

±2%

PLT

(0-99×109)/L

(±8.0×109)/L

100-999×109)/L

±10%


You May Like

Find Similar Products By Category

You May Like

View More
Chat Now Contact Now