Laboratory walk in test chamber testing temperature humidity climate cryo test chamber
Negotiable
Min.Order:1
Guangdong Huanrui Test Equipment Co., Ltd.
Walk-in Temperature and humidity test chamber use for industry large parts, semi-finish product,finish products
temperature and humidity testing. It is composed of control panel,switchboard,hydrating panel,fan, heater,humidifier,refrigerator.
Walk-in Temperature and humidity test chamber can create and keep temperature humidity constant space,and completely automatic control refrigeration,heating,dehumidify,humidification,supplementary humidified water.It can be added (optional) air exchange function,operator can long term work in it .
Meet the Standards
IEC60068-2-1.1990 testing methods for low-temperature
IEC60068-2-2.1974 testing methods for high-temperature
GJB150.
Character
Reference parameter
Model |
RHPW-4 |
RHPW-6 | RHPW-8 | RHPW-13 | RHPW-17 | RHPW-21 |
System | Balanced temperatuer & humidity control system | |||||
Testing temperature range |
-65℃~+120℃ |
|||||
Humidity range |
20%-95%RH |
|||||
Temperature fluctuation |
±0.5℃ |
|||||
Humidity fluctuation |
±2.5%RH |
|||||
Temperture uniformity | ±2.0℃ | |||||
Humidity uniformity |
±5.%RH |
|||||
Temperture deviation |
±2℃ |
|||||
Humidity deviation |
±3%RH (≥75%) ±5%RH(<75%) |
|||||
Heat up time |
20℃~+80℃ within 60 minutes | |||||
Pull down time |
20℃~-55℃,-40℃,-30℃,-20℃,-10℃ within 120 or 90 or 60 minutes. | |||||
Interior volume(m3) | 9.04 | 12.9 | 16.78 | 27.09 | 35.21 | 43.34 |
Interior material | SUS#304 Stainless steel plate | |||||
Exterior material | Model steel | |||||
Insulation material | Rigid polyerethane foam +glass fibre | |||||
Floor bearing load | 500Kg/m3 | |||||
Refrigeration system | Air cooled (Water cooled type is option) | |||||
Cascade type, hermetic/semitight compressor,CFC free refrigerant | ||||||
Controller | Color LCD screen in English/Chinese, programing,communicating function | |||||
Safety devices | No welding wire switch,compressor overheating and over current protection,over temperature protection, blower overload protection,dry heating machine protector,low water protection,fault warning system. | |||||
Power | AC380±10% 50HZ 3 phase 4 ires+ground wires |
Optional accessories