portable LED screen chemotherapy Animal Peristaltic iv medical syringe infusion pump for hospital
Negotiable
Min.Order:1
CE approved cold laser therapy fungus treatment Nail Fungus Laser Therapy Device
CE Approved Digital Electronic BP machine blood pressure monitor gauge with voice function
Wholesales Handheld High Image Quality Black and White Portable Ultrasound Machine
Medical Standard OEM Manual Aneroid Sphygmomanometer with Single or Dual Stethoscope
Hot selling handheld hospital house medical portable infrared vein finder/Detector/Vein viewer
Main features | Reliable and Safe: 1, Additional set of bubble alarm design(double bubble alarm) ensure more secure infusion; 2, Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can ues it safely; multipurpose: It can be transfused, transfused blood and transported nutrient solution. |
Pumping way | Semi-squeeze finger-type peristaltic pump |
IV-sets | Special or quality PVC/TPE infusion sets with an external diameter of 3.4mm~4.5mm |
User options | no |
Infusion modes | 4 infusion modes: 1, Flow-rate: ml/h; 2, Time-amount: ml/h; 3, Drop-rate: d/min; 4, Body-weight: mg/kg/h |
Parameter range | 1, Flow-rate: 1.0~1200.0 ml/h, 0.1ml/h step; 2, Time: 1s~99h59m59s, 1s step; 3, Drop-rate: 1~333 d/min, 1 d/min step; 4, Dose: 0.1~999.9 mg, 0.1mg step; 5, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step; 6, Body-weight: 1~300 kg, 1kg step; 7, KVO rate: 1.0~5.0 ml/h, 0.1 ml/h step; 8, Bolus rate: 50.0~800.0 ml/h, 0.1 ml/h step; 9, Bolus volume: 0.1~99.9 ml/h, 0.1ml step |
Precision | Better than ±5% (using a calibrated or a special IV-set) |
Alarm | 1, Air in tube, 5 grades of sensitivity (double bubble alarm); 2, Bubble sensor fault; 3.1, Upper-end occlusion, -20~ -60 kPa,8 grades of sensitivity; 3.2, Upper-end base-pressure fault; 4.1, Lower-end occlusion, 20~ 140 kPa,8 grades of sensitivity; 4.2, Lower-end base-pressure fault; 5.1, Pump rotation fault; 5.2, Inverse-infusion; 6, Abnormal door-opening; 7, Low battery power, battery fault; 8, Overtime pause; 9, Completed infusion; 10, AC power fault; 11, MCU fault, and stop infusion; 12, Battery temperature is too high. |
Data function | 1, Memory of calibration: Records the calibration data of tube automatically for application; 2, Memory of parameter: Records the latest infusion parameter for next operation; 3, Historical records: records latest 1000 infusion events. (what/when/why,flow-rate, etc.) |
Safety classification | Class I; Internal power supply, CF type, IPX4 equipment |
Power supply | AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery |
Power consumption:20VA | |
Working time of battery: 7h@25ml/h | |
Weight and size | 130mm(w)*206mm(h)*118mm(t); about 1.6kg |
Optional accessories | 1, Warmer; 2, Wireless communicating module. |