Coulometric Method Oxygen Permeability Testing Machine Flexible Packaging Lab Tester
Negotiable /Set
Min.Order:1 Set
Systester Instruments Co., Ltd
Separator Gas Permeate Test Air Permeability Tester Gurley Lab Test
12 Tests In One Univerisal Tensile Tester For Rubber Sheet/ Plastic Membrane
Max Gas Permeability Testing Equipment EVR System Separatiom Membrane
Gurley Test Method Air Permeability Tester For Paper/ Membrane
Vacuum Leak Test Procedure Of Flexible Packaging Lab Leakage Testing Machine
Coulometric Method Oxygen Permeability Testing Machine Flexible Packaging Lab Tester
Product Details
Application:
Based on the coulometric sensor method and available for the OTR test of flexible films, sheeting and packages. It can test 3 specimens in one operation with independent results.
Films:
Plastic films, composite films and many others.
Packages:
Plastic packages.
Coulometric Method Oxygen Permeability Testing Machine Flexible Packaging Lab Tester
Features:
— Coulometric sensor method
— Own technology
— Film and package test
— 3 specimens test simultaneously and independently
— One-button test, test automatically in whole process
— Multiple test modes optional
— Data curves display
— Computer control
— Digital flow adjustment
— 3 Chambers temperature sensor embedded in
— Environment temperature and humidity sensor inside
— Chambers water bath technology
— Unit based on embedded system,24bit Δ-Σ AD
— Pressure, oxygen and temperature overload protection
— Traceable reference film calibration
— Supports DSM system (DSM, lab data management system)
Principles
Film:
The specimen is clamped between the upper and lower chamber, the upper chamber is purged by a oxygen stream, and the other side of specimen, lower chamber is purged by a stream of nitrogen, due to the effect of oxygen concentration gradient at two sides of the specimen, oxygen molecules will permeate through the specimen into the nitrogen side, penetrated oxygen molecules will be carried to the coulometric sensor which generates the proportional signal, so oxygen transmission rate and permeability coefficient will be obtained by calculating the sensor signal.
Package:
The package is mounted on the lower chamber, the outside of the package is air or a pure oxygen stream, the inside of the package is purged by a nitrogen stream, oxygen will permeate through the wall into the inside of the package, penetrated oxygen molecules will be carried to the coulometric sensor which generates the proportional signal, so oxygen transmission rate will be obtained by calculating the sensor signal.
Coulometric Method Oxygen Permeability Testing Machine Flexible Packaging Lab Tester
Technical indexes:
Test range: | |
Film | 0.01-6500 cm3/m2·d (Standard) |
0.07-6,5000 cm3/m2·d (Optional) | |
Resolution | 0.001 cm3/m2·d |
Package | 0.0001~62 cm3/pkg·d |
Resolution | 0.00001 cm3/m2·d |
Specimen amount | 3 specimens, independent |
Temperature range | 5C – 95C |
Temperature accuracy | ±0.1C |
Humidity range | 0%RH,35%RH-90%RH,100%RH |
Humidity accuracy | ±1%RH |
Test area | 50cm2 |
Specimen size | |
Film | size >150*94mm |
Thickness | less 5mm |
Package | |
Max size | D120mm * 400mm |
Carrier gas | 99.999% Nitrogen |
Carrier flow | 0~20 ml/min |
Gas interface | 1/8 inch |