Cheap High Sensitive Gas Chromatograph Price
USD $3,000 - $15,000 /Piece
Min.Order:1 Piece
Shanghai Drawell Scientific Instrument Co., Ltd.
DW-GC1120 High sensitivity gas chromatograph
Temperature Control :
Temperature Range:7℃ ~ 400℃ above room temperature, (increment 1℃)
Temperature Control Object:Column Box, front and back detector, front and back sampler, assist,total six channels.
Program ramps : 7 ramps in total
Program setting:0.1℃~40℃/min(increment 0.1℃)
Total time of total program : Up to 655 minutes, (increment 0.1℃)
Hydrogen flame ionization detector (FID):
Minimum detectable quantity: M≤8×10-12g/s (C16)
Best test result: M≤3×10-12g/s (C16)
Noise: ≤5×10-14A
Drift: ≤6×10-13A/h
Linearity range: ≥106
Thermal conductivity detector(TCD):
Sensitivity:≥5000mV.mL/mg (C16)
Noise: ≤20µV
Drif: ≤30µV/h
Linearity rang: ≥104
High Sensitivity Thermal conductivity detector(HTCD):
Sensitivity: ≥10000mV.mL/mg (C16)
Electron capture detector(ECD):
Detectability: ≤1×10-13g/s (γ-666)
Linearity range: ≥103
Maximum using temperature: 350℃
Flame photometric detector(FPD):
Detectability: ≤2×10-11g/s(P) (Parathion-methyl)
≤1×10-10g/s(S) (Parathion-methyl)
Nitrogen-phosphorus detector(NPD):
Detectability: ≤5×10-12g/s(N) (Azobenzene)
≤5×10-13g/s(P) (Malathion)
(On programmed)
Photoelectric ion detector(PID):
Detectability: ≤1ppb (Benzene)
Optional Parts | Main function |
GC1120-TCD | Thermal conductivity detector |
GC1120-ECD | Electron capture detector |
GC1120-FPD | Flame photometer detector |
GC1120-NPD | Nitrogen phosphorous detector |
Chromatographic work station | External interface box + work station (not include PC) |
Simulation software | Imitation find chromatographic condition, emulation teaching, train and knowledge base. |
Instrument technical index
Flexible Sample Introduction System
Up to three injectors may be installed and operated simultaneously with independent temperature control. Wide range of injectors is available to be chosen from for various applications. It is very convenient for installation and disassembly of sample injectors.
Sample Injector with Mono Packed Column/Double Packed Columns:
Several alternative sample introduction methods, including on-column sample introduction and six-way valve injection;
Adapter for injection to capillary column (0.53 mm i.d.).
Sample injector for capillary column
Special sample injector with septum purge and back-pressure valve splitting;
Special programmable split/splitless injector with solenoid valve, septum purge, back-pressure valve splitting, and standalone carrier gas and make-up gas flow;
Modified dual-packed-column sample injector for injection to capillary column by employing some connectors and back-pressure valve splitting.
Advanced Pneumatics
Carrier gas control with constant pressure valve and constant flow valve;
Air and hydrogen control with constant pressure valve and needle valve;
Flow easily adjustable through constant flow valve and needle valve with digital display;
High precision, reproducibility and reliability.
Sensitive and reliable detector
Up to three detectors (FID, TCD, ECD, etc) may be installed and operated simultaneously. Simultaneous multi-detection with detectors in series or in parallel meets the various analysis requirements.
The new FID provides higher sensitivity than ever before. It is very convenient to disassemble and clean the jet and collector. Precisely fixed emitter configuration ensures conformity of product performance;
Excellent TCD detector with high-resistance tungsten-rhenium filaments provides high sensitivity. A sensitivity of 10000 mV·mL/mg is achieved with optional amplifier circuit board. Special configuration of air heat insulation enhances the stability of TCD detector
Constant-current, pulse-modulated ECD detector sets sensitivity and current by computer.
Advanced Microcomputer Control System
With advanced control techniques, microcomputer-based temperature control system provides superior performance.
High temperature precision (± 0.5℃), high reliability, and high resistance to disturbance;
Five independent heated zones, with maximum operating temperature of 400℃;
Temperature-limit setting and overheat protection.
Instrument control is accomplished through the keypad, and the instrument offers outstanding capabilities.
Self-diagnosis;
Power-off protection;
Ten methods saving and reloading;
Two controls of external events;
Timing on/off of temperature control;
Timing program, detector status, measurement range, polarity and current setting and display;
Temperature setting values, actual values, retention time and analysis time display.
High-performance spacious column oven
Spacious column oven accommodating capillary column and dual packed columns and convenient for column installation;
Fast heat-up and rapid cool-down column oven with automated back-door opening (300℃ to 50℃ in 7 min);
Available for both near-ambient and above ambient temperature control;
Supports 5 oven ramps with 6 plateaus.