DTR2000 Portable Raman Spectrometer
FEATURES
Ø linear CCD with TEC cooling;
Ø low noise circuit;
Ø Powerful embedded software;
Ø Fluorescent background eliminate;
Ø Peak searching and indicating;
Ø 10.1’’ LCD;
Ø Android Operation System;
Ø USB 2.0;
Ø Touch screen and Keyboard input;
Ø Friendly HMI;
Ø Battery Endurance > 5 hours;
Ø Remote control via LAN;
Ø IP67 case;
APPLICATIONS
Ø Biological science
Ø Pharmaceutical engineering
Ø Forensic analysis
Ø Agriculture and food safety
Ø Gemstone
Ø Environmental science
GENERAL DESCRIPTION
DTR2000 portable Raman spectrometer is suitable for field operation. The outstanding reliability makes the detection result much accurate. The excellent low stray light conditions that enable the spectrometer has a wide range of application, especially in the biochemical analyzer, food safety, pharmaceutical engineering. The multi-function software facilitated the spectral analysis process in application. The remote experiment through internet access makes the bald test item much easier.
1 Specifications
DTR2000 System | |
Interface | USB 2.0/LAN |
OS | Android 4.4.2 |
Battery endurance | >5 h |
Integration time | 4ms - 120s |
Supply voltage | DC 19V(+/-5%) |
Operating temperature | 5~40 ℃ |
Operating humidity | < 95% |
Size (L*W*H) | 30×22.5×13.2 cm3 |
Weight | 7 Kg |
Reliability | |
Spectral stability | σ/μ < 0.5% (COT 8 hours) |
Temperature stability | Spectral shift ≤ 1 cm-1 (10-40 ℃) |
Variation of Spectral intensity in 5 ~ 40 ℃ | <±5% |
Optics Parameter | |
Spectral region | 250-2400 cm-1 |
Spectral resolution | 6 cm-1 |
Signal-to-Noise | >200:1 (918 cm-1 of Acetonitrile,10s accumulation, 200mW) |
Incidence slit | 50 μm |
Optical system | f/4 Crossed Symmetrical Czerny-Turner |
Focal length | 98 mm for incidence and output |
Detector | |
Type | Linear Array Detector |
Detectable range | 200-1100 nm |
Effective pixel | 3648 |
Dynamic range | 1300:1 |
Pixel dimension | 8μm×200μm |
Full well capacity | 100 Ke- |
Sensitivity | 130 Photon @ 400 nm; 60 Photon @ 600 nm |
Exciting Laser | |
Center Wavelength | 785nm (+/-1nm) |
FWHM | 0.1 nm |
Output power | ≥450 mW |
Power stability | σ/μ <±0.2% |
Raman Probe | |
Working distance | 6 mm |
Rayleigh scattering resistance | OD>8 |
Numerical Aperture | 0.3 |
Aperture | 7mm |
2 Optical Performance
2.1 General spectral performance
2.2 Spectral Resolution
2.2.1 Raman spectral of Tylenol
2.2.2 Raman spectral of petrol
.
3 Reliability
Figure 3.1 and Figure 3.2 showed the temperature reliability testing results of fives DTR2000 portable Raman spectrometers. The testing temperature range was from 5 oC to 40 oC. The spectrometer was kept more than 1 hour at every temperature spots. Acetonitrile was used as the standard sample in the testing. The testing results were calculated using 918 cm-1 of acetonitrile. The wavenumber shift was 1 cm-1 or less(as show in Fig. 3.1). The peak intensity variation was less than 10% (as show in Fig. 4).
You May Like
- DTR3010 Portable handheld Cooled raman spectrometer portable
- DTR2000 Portable Raman Spectrometer Prices China Drawell
- raman handheld spectrometer price
- DW-CS-8820 High-frequency Infrared Carbon Sulfur Analyzer
- DW-CS-8620 Arc Infrared carbon and sulfur analyzer
- DW-CS-8620 Arc Infrared Carbon&Sulphur Analyzer Price Drawell
- DW-CS-8820 Series Drawell High-frequency Carbon&Sulfur Analyzer
- DW-NP-5010 Energy Dispersion X-ray Fluorescence Spectrometer XRF
- DW-TY-9000 Direct Reading Spectrometer for Metal Analysis
- Metal Analysis Direct Reading Spectrometer Prices for DW-TY-9000
Popular Searches
Find Similar Products By Category
You May Like
- DTR2000 Portable Raman Spectrometer with Probe
- DTR3010 Drawell Laboratory Portable Raman Spectrometer Prices
- DTR3010 Portable Raman Spectrometer
- DW-AA320N Lab Atomic Absorption Spectrophotometer Price
- AA4530F Single Beam Atomic Absorption Spectrophotometer
- Hollow Cathode Lamp for AAS Atomic Absorption Spectrophotometer
- Flame AAS Atomic Absorption Spectrophotometer DW-AA4530F
- Flame AA320N Atomic Absorption Spectrophotometer
- Graphite Furnace Atomic Absorption Spectrometer Price
- Flame Atomic Absorption Spectrometer Metal Analyzer
- Drawell AAS Double Beam Atomic Absorption Spectrophotometer
- Drawell Infrared Carbon Sulfur Analyzer
- Arc Infrared Carbon Sulfur Analysis Instrument DW-CS-8620
- CS Infrared Carbon Sulfur Analyzer DW-CS-8620
- Full Spectrum Direct Read Spectrometer with high quality