High and low temperature environment simulation test chamber with temperature range of -120 degrees to 150 degrees
USD $3,000 - $15,000 /Set
Min.Order:1 Set
high and low temperature cycle test temperature probe testing equipment electric oven for Laboratory
High and low temperature test chamber environmental test equipment
High Low temperature test chamber environmental test equipment with CE certificate
environmental climate high and low temperature test chamber manufacturer
Product Show: |
Product Advantages: |
Product Process: |
Product Parameters: |
High and low temperature test chamber
Product Usage:
High and low temperature test chambers are widely used in aviation, aerospace, electronics, instrumentation, electrical products, materials, automotive parts, plastics and rubber products, chemicals, building materials, medical, photovoltaic and other industries.They test the temperature stress, temperature aging screening, reliability testing, performance testing, weathering test, high temperature storage of test pieces under high temperature and environmental conditions.
Implementation and meeting standards:
GB/T 10589-2008: cryogenic test chamber technical conditions
GB/T 10592-2008: high and low temperature test chamber technical conditions
GB/T 2423.1-2008: low temperature test method
GB/T 2423.2-2008 : high temperature test method
GB/T 2423.22-2008: temperature change rate test method
GJB 150.3A-2009: high temperature test method
GJB 150.4A-2009: low temperature test method
GJB 360B-2009: test methods for electronic and electrical components
JJF 1101-2003: environmental and test equipment temperature and humidity calibration specifications
GB/T 5170.2-2008: test method for environmental test equipment of electric and electronic products temperature test equipment
Product features:
New type of refrigeration: compressor refrigeration can achieve any temperature from -70°C to -170°C
Efficient and energy-saving: reduce power consumption by more than 30% during low temperature test
Excellent performance: performance indicators better than national and military standard requirements
Compact size: easy access to the laboratory within 500 liters, saving floor space
Optimize the air duct: based on the professional CFD fluid mechanics design, the air circulation is more reasonable
Simple operation: user-friendly operation interface, one-stop start-up, no professional training
Multiple Protection: ensure the safety of personnel, test pieces and equipment, worry-free use
Technical indicators:
product name | High and low temperature test chamber | ||||
Product number | MQ-T100F-2 | MQ-T250F-2 | MQ-T500F-2 | MQ-T1000F-2 | |
Nominal volume | unit/L | 100 | 250 | 500 | 1000 |
Inner size | Width × depth × height (mm) | 450×450×500 | 600×600×700 | 750×750×900 | 1000×1000×1000 |
External size | Width × depth × height (mm) | 650×1000×1450 | 800×1200×1700 | 950×1400×1900 | 1200×1500×1950 |
Performance | Temperature range | A,-40℃~150℃ B,-70℃~150℃ C,-90℃~150℃ D,-120℃~150℃ | |||
Temperature fluctuation | ±0.5℃ | ||||
Temperature uniformity | ≤2.0℃ | ||||
Temperature deviation | ≤±2.0℃ | ||||
Heating rate | ≥2.0℃/ Min | ||||
Cooling rate | ≥2.0℃/ Min | ||||
Linear temperature control | 0.5~1.0℃/Min | ||||
Refrigeration method | Compressor refrigeration | ||||
Cooling method | Air-cooled | ||||
The main components | Control System | Siemens programmable controller + touch screen + independent development program | |||
Heating system | Fully independent system, high efficiency stainless steel finned tube heater | ||||
Cooling System | Importer-type scroll compressor | ||||
Circulatory system | Temperature and low noise motor, stainless steel centrifugal impeller | ||||
Use materials | Carton material | High-quality carbon steel plate, phosphating electrostatic spray treatment | |||
Inner box material | International standard SUS304# stainless steel mirror treatment | ||||
Insulation material | Hard flame retardant polyurethane foam | ||||
Sealing material | High and low temperature environment-friendly silicone rubber seal | ||||
voltage | 380±10% VAC,50Hz,3L+N+G | ||||
Communication Interface | RS232 communication interface, USB data interface | ||||
Standard configuration | Sample holder 1 set, test lead hole 1 set | ||||
Optional accessories | Recorder, Remote Control Software, Insulated Sample Holder, Split Manipulator, Dry Air Purge | ||||
safety devices | Leakage, short circuit, motor overheating, heater overheating, compressor overpressure, overload, multiple overtemperature protection | ||||
Environmental requirements | Ambient temperature: 5~35°C, relative humidity: ≤85%RH, no strong vibration, electromagnetic radiation, no dust and corrosive substances around | ||||
Note: 1.temperature performance parameters are in the ambient temperature of 25 °C, studio no load conditions, in accordance with GB/T5170.2-2008 test; 2. according to the specific requirements of users customized non-standard high and low temperature test box; 3. This technical information is subject to change without notice. |
High and low temperature test chamber common fault solution
1, the temperature drops below the preset value or can not cool down
(1) The compressor is damaged or not opened, replace the compressor and check the compressor circuit and control components for damage.
(2) Refrigerant leakage, use nitrogen leak detection method to find the leak point, weld, vacuum, add refrigerant, re-commission
(3) Solenoid valve is abnormal, replace the solenoid valve, check the solenoid valve control circuit
(4) Capillary clogging, replacing the capillary, replacing the dry filter, blowing the condensing system impurities with nitrogen, adding refrigerant debugging
(5) The condenser freezes, heats up the moisture in the tank, and then cools down.
2, the temperature rises below the preset value or can not heat up
(1) The heating tube is damaged, and the heating tube is replaced.
(2) Solid state relay is damaged, replace solid state relay
3, the humidity does not reach the preset value or can not humidify
(1) The humidification pipe is damaged, and the humidification pipe is replaced.
(2) The water level float that controls the humidification does not float
(3) Humidification solid state relay is damaged, replace the humidification solid state relay
4, humidity always shows 100% RH
(1) Abnormal gauze, change gauze
(2) Failure of the refrigeration system, check the refrigeration system
(3) Abnormal water supply in the waterway system, check the water system
5, the main power switch trip
(1) The total power of the equipment exceeds the total current of the power switch, and the appropriate power switch is replaced.
(2) Leakage of equipment, check compressor, motor, heating pipe, humidification pipe, condenser, fan, transformer, switching power supply, water pump and other components, insulation resistance and leakage current
6, leaking water
(1) Check whether the water pipe, water cup, humidifying bucket and water tank are normal. If there is any abnormality, use the raw material tape for inspection or seal with high temperature resistant neutral glass glue.
(2) Check whether the seal inside the box is normal, if there is abnormality, seal with high temperature resistant neutral glass glue.
(3) Check whether the refrigeration system is normal. If there is any abnormality, use insulation cotton to wrap the insulation.
7. Frost or ice in the box
(1) Long-term operation of low temperature causes the temperature to cool after drying at high temperature
(2) When the test is carried out, the door is often opened, and the door is not opened or opened as much as possible during the test.
(3) The circulating motor is damaged or the wind wheel is removed, the circulating wind wheel is replaced and the wind wheel is tightened.
8, the equipment is noisy
Detect the motor, wind wheel, condenser, fan, compressor, casing cover, etc.
Related Products: |
Click the pictures for detail product info
Packing & Delivery: |
About Us: |
Zhongkemeiqi (Tianjin) Technology Co., Ltd, which is a high-tech enterprise specializing in the research, development, production and sales of many kinds’ climatic chambers.
We have passed the ISO9001 quality management, and our team consists of several senior experts who have been engaged in the development, manufacture and maintenance of environmental test equipment for many years.
Temperature and Humidity test chamber,
Rapid Temperature Change test chamber,
Thermal Shock test chamber,
Temperature and Low-Pressure test chamber,
Temperature Humidity and Vibration test chamber;
Ultra-Low Temperature Chamber /Solution,
Aging test chamber (UV, Xenon lamp, Ozone) ,
Rain / Sand Dust test chamber,
Vacuum drying oven, …
Factory Display: |
Our Certificates: |
FAQ: |
Q1: How can i get a quotation?
A1: Kindly provide us your details request (inner chamber size, temperature range, humidity range, power supply, product, etc.), leave us an inquiry or email and we will response promptly!
Q2: What's your temperature & humidity range?
A1: Our standards temperature range is -70℃~+180℃, 20%~98%RH.
We also can do Ultra low temperature to -190℃ .
Q3: What's your rate of heating & cooling?
A3: Our standards rate is average 3℃/min for heating, 2℃/min for cooling.
3℃/min,5℃/min,8℃/min,10℃/min,15℃/min linear or non-linear speed is available for us.
Q4: What's your warranty?
A4: 12 Months (Noted: free spares parts can be offered during warranty period, exclude consumables and man-made damages), life-time technical service
Contact Us: |