Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10
Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10
Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10
Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10
Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10
Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10

Acrel 300286 medical IT intelligent insulation monitoring device AIM-M10

USD $1,243.9 - $1,243.9 /Piece

Min.Order:1 Piece

Supply Ability:
1000 Piece / Pieces per Month
Payment Terms:
T/T Credit Card

Quick Details View All >

Brand Name:
Acrel
Place of Origin:
Jiangsu, China
Model Number:
AIM-M10
Operating Temperature:
-10~55℃
Dimensions:
54*85*63.6mm
Display Type:
Digital Only

Jiangsu Acrel Electrical Manufacturing CO.,LTD

Business license
Business Type: Manufacturer
Wuxi Jiangsu China
Main Products: power sensor ,programmable power meter ,DC power analyzer ,Data center monitor device ,Din rail energy meter ,Isolated power supply ,wireless temperature sensor ,IOT energy meter ,split core current transformer

Product Details

Product Description

General

Hospital IPS monitor system insulation monitor AIM-M10 with CE approval

medical intelligent insulation monitoring instrument adopts advanced microcontroller technology, which has high integration,
compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is ideal selection for
insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.

Features

* real-time monitoring and fault alarming of the ground insulation resistance,
* transformer load current and transformer winding temperature of the monitored IT system,
* be used with insulation fault locator, remotely starting fault-locating and displaying locating results when there are
insulation faults,
* real-time monitor the line disconnection fault, temperature sensor disconnection fault and the functional grounding line
disconnection fault of the monitored system, and give the alarm indication within 2S after the fault occurs,
* relay alarm output, LED alarm indication and other faults indication functions,
* used for centralized alarm and display instrument, test signal generator, insulation fault locator and upper computer management
software communications, and can monitor the operation status of IT system in real time.
* Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to
analyze the operation conditions of system and promptly eliminate the faults.

Dimension

Wiring

Typical connnection

Techinical Parameter

Model
AIM-M10
Voltage
AC220V( fluctuation range ±10%)
Frequency
50/60Hz
Max power consumption
<8VA
Insulation resistance measuring range
50~999kΩ
Relative percentage accuracy
0~±10%
Alarm value range
10~999kΩ
Response time
<2S
Measuring voltage
<12V
Measuring current
<42μA
Measuring range
2.1~50A
Alarm value range
5~50A
Measuring accuracy
class 5
Display
Digital and percentage display
thermal resistor
PT100
Measuring range
-50℃~+200℃'
Alarm value range
0℃~+200℃'
Output type
1 channel relay output(unable to program)
Contact rating
AC 250V/3A DC30V/3A
Working temperature
-10℃~+55℃'
Storage temperature
-20℃+70℃'
Relative humidity
5%~95%,non-condensing
Altitude
≤2500m
Communication
one channel RS485,MODBUS-RTU
Rated impulse voltage/pollution class
4KV/III
EMC and electromagnetic radiation
In accordance with IEC1326-2-4

Certificate

Contact Supplier

Mr. Tim Overseas sales engin Chat Now
Telephone
86-1870-6168051
Mobile
86-18761599716
Fax
86-1870-86179975
Skype
86 18761599716
Address
Jiangyin Wuxi,Jiangsu
Whatsapp
86 18761599716

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Hi~