TEJS-6000 Automatic dielectric loss test kit
USD $1999 - $3999 /Piece
Min.Order:1 Piece
Baoding Tianteng Electric Co., Ltd.
Automatic Dielectric Loss Tester
Model: TEJS-6000
Product Description: Dielectric loss measurement is a basic method in insulation test, degradation and partial defect of insulation of electrical equipment. It adopts high-power switching power supply, which outputs 45Hz and 55Hz pure sine wave and automatically increase the voltage to a maximum value 10 KV. It can filter 50Hz interference automatically, and it is applicable to the site test of substation and other places with large electromagnetic interference.
Technical data:
Accuracy: Cx:±(reading×1%+1pF)
tgδ: ±(reading×1%+0.00040)
Anti-interference: Variable frequency anti-interference, the above accuracy can be reached under 200% interference.
Capacitance range: Internal HV: 3pF~60000pF/10kV 60pF~1μF/0.5kV
External HV: 3pF~1.5μF/10kV 60pF~30μF/0.5kV
Resolution: Maximum 0.001pF, 4 valid figures.
tgδ range: unlimited, 0.001% resolution, automatic identification for capacitance, inductance and resistance of three tested articles.
Test current range: 10μA~1A
Internal HV: Set voltage range: 0.5~10kV
Maximum output current: 200mA
Buck-bust method: continuous stepless regulation
Test frequency: 45, 50, 55 single frequency
45/55Hz automatic dual variable frequency
Frequency accuracy: ±0.01Hz
External HV: In case of UST, maximum test current is 1A, and power frequency or variable frequency is 40-70Hz
In case of GST, maximum test current is 10kV/1A, and power frequency or variable frequency is 40-70Hz
CVT self-exciting low voltage output: Output voltage 3~50V, output current 3~30A
Measuring duration: About 40s, depending on measuring method
Input power supply: 180V~270VAC, 50Hz/60Hz±1%, supplied by alternating current or generator
Computer interface: Standard RS232 interface
Printer: WH-A7 mini-type thermal printer