use method
1.After open the package, check it whether is intact, and the furnace should be placed on the flat ground , protect the temperature controller from vibration.
2.Install the power switch to control the power supply. In order to ensure safe operation , the furnace and temperature controller should be connected with protective conductor thermal
3.Check the wire connection, close lid again. then connect the power, adjust the temperature adjustor to “set”, set needed temperature. When it displays needed temperature, Green light is on, it means that heating starts .then dial to “measure”, it displays the actual temperature. and ammeter displays the corresponding value according to the power (8-10). meanwhile the temperature rises, when the temperature reaches setting value, controller will disconnect power supply automatically, red light is on or ammeter in the position of “0”, it means that heating stops. then “red” and “green” light of temperature controller will work alternately, to reach constant temperature
Note:
When equipment is running for the first time, the temperature will soar(red light is on or ammeter in the position of “0”, but temperature keeps rising), but will be back later
Safety Notice
1. When this furnace is running the first time or set it aside for a long period.
2. When using this furnace, furnace temperature must be lower than rated temperature, or it will damage heating parts. and avoiding directly pour into any liquid or dissolved metal .please frequently remove scraps or oxide inside the furnace to keep it clean.
3. The inside furnace temperature has a lagged effect on the temperature detected from temperature sensor because of large heating power, and temperature will soar as the change of ambient temperature .then controller will adjust temperature automatically until the displayed temperature is the same as setting temperature. If user need to reduce temperature soar , first step: if the needed temperature is 500℃,then set temperature at 400℃,when temperature goes up to 400℃, set temperature up to 450℃ again, when it reaches 450℃,then set it up to 500℃ .
4. when using this equipment, must follow bellows:
1)Altitude is less than 1000 meter
2)Ambient temperature is +5-40℃
3)The mix monthly average humidity is less than 90% , at the same time the minimum monthly average temperature is less than 25℃.
4)There is no conductive dirt explosive gas or corrosive gas around the furnace
5)Place this furnace on the flat and firm ground