Building Material Water Vapor Permeability Test ASTM Standard Lab Testing Equipment
Negotiable /Set
Min.Order:1 Set
Systester Instruments Co., Ltd
Li-ion Battery Separator Membrane Air Permeability Testing Machine Gurley Method
Explosive Gas Permeability Test High Pressure Gas Permeability Lab Testing Machine
Green Energy Separator Membrane Air Permeability Testing Machine Gurley Standard
Auto Servo Tensile Testing Machine Pharmaceutical Packaging Elongation/ Puncture/ Tension Test
Plastic Membrane/ Flexible Packaging Water Vapor Permeability Lab Testing Apparauts
ASTM Gas Permeability Test Barrier Material Permeate Testing Machine
Building Material Water Vapor Permeability Test ASTM Standard Lab Testing Equipment
Products Details
Application:
Based on the electrolytic sensor method and available for the water vapor permeability test of flexible films, sheeting and packages. It can test 3 specimens in one operation with independent results.
Films:
Plastic films, composite films and many others.
Packages:
Plastic packages.
Building Material Water Vapor Permeability Test ASTM Standard Lab Testing Equipment
Features:
— Electrolytic sensor method
— Own technology
— Film and package test
— 3 specimens test simultaneously and independently
— One-button test, test automatically in whole process
— Multiple test modes optional
— Data curves display
— Computer control
— Digital flow adjustment
— 3 Chambers with 3 temperature sensors
— Environment temperature and humidity sensor inside
— Chambers water bath technology
— Unit based on embedded system,24bit Δ-Σ AD
— Pressure, ppm water sensor and temperature overload protection
— Traceable reference film calibration
— Supports DSM system (DSM, lab data management system)
Building Material Water Vapor Permeability Test ASTM Standard Lab Testing Equipment
Technical indexes:
Test range: | |
Film | 0.001-52 g/m2·24h (Standard) |
0.01-1100 g/m2·24h (Optional) | |
Resolution | 0.001 g/m2·24h |
Package | (optional) |
0.0001~0.5 g/pkg·d | |
Resolution | 0.00001g/pkg·d |
Specimen amount | 3 specimens, independent |
Temperature range | 5C – 95C |
Temperature accuracy | ±0.1C |
Humidity range | 0%RH,35%RH-90%RH,100%RH, standard condition 90%RH |
Humidity accuracy | ±1%RH |
Test area | 50 cm2 |
Specimen size | |
Film | size |
Thickness | less 5mm |
Package | |
Max size | D120mm * 400mm |
Carrier gas | 99.999% Nitrogen |
Carrier flow | 0~200 ml/min |
Gas interface | 1/8inch |