Ifrared/ Weighing Method Water Vapor Permeability Testing Machine Medical Golves Permeate
Negotiable /Set
Min.Order:1 Set
Systester Instruments Co., Ltd
Contact Lenses/Keratoscope Oxygen Permeability Testing Machine ASTM Standard
Opening Force Of Plastic Bottle Cap Torque Testing Machine Medical Bottle/ Pharmaceutical Packaging
ASTM D4964 Elasticity Test Of Fabrics Universal Testing Machine Servo Tensile Tester
Medical Infusion Bags And Films Water Vapor Permeability Test Lab Testing Apparatus
Degradable Packaging/ Compostable Materials Gas Permeability Lab Testing Machine Food Package
Standard Gurley Method Seperator Membrane Performance Air Permeability Testing Machine
Ifrared/ Weighing Method Water Vapor Permeability Testing Machine Medical Golves Permeate
Product Details
Application:
Based on the humidity sensor detection method and available for the water vapor permeability test of heat exchange membranes, nonwoven fabrics and flexible barrier materials.
Ifrared/ Weighing Method Water Vapor Permeability Testing Machine Medical Golves Permeate
Features:
features | Weighing method |
Dual weighing modules, own technology | |
18 specimens, continuously test | |
One-button test, test automatically in whole process | |
Multiple test modes optional | |
Temperature and humidity auto control | |
Data curves display | |
Computer control | |
Environment temperature and humidity sensor inside | |
Unit based on embedded system,24bit Δ-Σ AD | |
Traceable reference film calibration | |
Supports DSM system (DSM, lab data management system) |
Principles:
The specimen is clamped between the upper and lower chamber, the upper chamber is high humidity side, and the other side of specimen, lower chamber is purged by a stream of dry gas, due to the effect of water concentration gradient at two sides of the specimen, water molecules will permeate through the specimen into the dry gas side, penetrated water molecules will be carried to the relative humidity sensor which generates the proportional signal, so water vapor transmission rate will be obtained by calculating the sensor signal.
Ifrared/ Weighing Method Water Vapor Permeability Testing Machine Medical Golves Permeate
Technical indexes:
Test range | 5 - 6500 g/m2·24h (Standard) |
Resolution | 0.001 g/m2·24h |
Specimen amount | 1 specimen |
Temperature range | 5C–95C |
Temperature accuracy | ±0.1C |
Humidity range | 100%RH (Standard) |
Humidity accuracy | ±1%RH |
Test area | 38 cm2 (Standard) |
Specimen size | >=150mm * 94mm |
Thickness | less 3mm |
Carrier gas | Dry Air, or 99.999% Nitrogen |
Carrier flow | 0~200 ml/min |
Gas interface | 1/8inch |