Product Details

SH105C automatic low temperature closed cup flash point tester

Brand Name Shengtai Instrument
Place of Origin China
Model Number SH105C automatic low temperature closed cup flash point tester

Product Features

  

SH105C automatic low temperature closed cup flash point tester is designed and manufactured as per standards GB/T261Determination of Flash Point - Pesky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products in the temperature range from 40℃ to 360℃

Technical features

1. It adopts LCD screen to display.  It can preset the parameters of expected flash point,sample mark number, atmospheric pressure, test date,etc.It has menu to prompt and input function of guide type.

2. It adopts simulation tracking display the function curve of temperature rising and test time.It has prompt functions  mis-operation,test date and test time.

3. It is equipped with standard RS-323,485 computer port.The lower computer can save 100 groups of historical data. The test data can be saved for a long time, transmitted and modified if the instrument being connected with computer.

4. It can correct the influence of atmospheric pressure and calculate the correction automatically.

5. Differential detection.Automatically correct the systematic deviation.

6. Automatically open the lid,ignite,detect and print test data.The test arm lifting up and down automatically.

7. It adopts electric ignition to fire light the gas flame.It will light only need to press the button.

8. Humanized design,beautiful and safe. Easy to operate.

9. Cryogenic cycle device: compressor refrigeration

Technical specifications and parameters

1. Power AC(220±10%)V50Hz

2. Temperature measurementRange: 0300℃

Repeatability≤3℃

Reproducibility≤6℃

Accuracy0.1℃

3. Heating rateProcedure A(56)℃/min

Procedure B(11.6)℃/min

Automatic control and manually adjustable.

4. Stirring rate(1) Procedure A: (90120)RPM

(2) Procedure B: (250±10)RPM

Automatic control and manually adjustable.

5. Igniting modeElectric ignition. Gas flame diameter:3.2mm4.8mm

6. Working conditionambient temperature(1040)℃

relative humidity≤80%

7. Total power ≤500W

8. Overall dimension520mm×360mm×310mm 