Medical Fetal Multi-Parameter Patient Monitor 6000b
USD $1 - $999 /Set
Min.Order:1 Set
MERMAID (MEDICAL)INDUSTRY LIMITED
High Quality Clinic Patient Use ECG Machine, Multi-Parameter Patient Monitor 9000
Biphasic Portable Cardiac Defibrillator Me9000b Patient Monitor Defibrillator
Medical Deice Hospital Medical ECG Fetal Multi-Parameter Patient Monitor 9000d
Clinic Medical Patient High Quality Me9 Multi-Parameter Monitor with Ce/ISO Approved
Hospital Medical 12.1\'color Me6 Touch Screen Multi- Parameter Patient Monitor
Surgical Instrument Skin Care Rife High Frequency Welding Machine Me-50A
Medical Equipment Multi-parameter Patient Monitor 6000B
Multi-parameter Patient Monitor 6000B
7"color TFT screen
Multichannel display Big font interface
ST and arrhythmia analysis
Drug dose calculation and titration table
OxyCRG dynamic view display
Respiration monitoring with apnea alarms
Alarm review, arrhythmia review
Patient info. Input
Antidefibrillation design
NIBP over pressure protection
Li-ion batteries, detachable and rechargeable,
Continuous working 90 minutes;
AC/DC adaptable: 12V3A 100V-240V
Supporting wire network.
Type: Asystole, ventricular, fibrillation, ventricular tachycardia, paper non-capture, ventricular rhythm, couplet, VT>2, Bigeminy,
Trigeminy, Ron TPVC, Multiform PVC, irregular ,Rhythm, missed beats, bradycardia, tachycardia
Technical Specifications
ECG
Lead Mode 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, AVR, avL, avF, V
Waveform 2 CH
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 CH
Gain 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, auto
HR and Alarm Range Adult 15 ~ 300 bpm
Neo/PED 15 ~ 350 bpm
Accuracy ± 1% or1 bpm, which is greater 1 bpm
ResolutionSensitivity > 200 (UV PP)
Differential Input Impedance > 5 M
CMRR Monitor >105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential ± 300mV
Leakage Current<10 uA
Baseline Recovery<3 S After
Defi. ECG Signal Range ± 8 mV ( Vpp )
Bandwidth Surgery 1 ~ 20 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 300 Hz
Calibration Signal 1 (mV PP), Accuracy 5%
ST Segment Monitoring Range Measure and Alarm 2.0 ~ +2.0 mV
ARR Detecting Type: Asystole, ventricular, fibrillation, ventricular tachycardia, pacer non-capture, ventricular rhythm, couplet, VT> 2, Bigeminy, trigeminy, Ron T PVC, Multiform PVC, irregular rhythm, missed beats, bradycardia, tachycardia
Respiration
Method Impedance between RF(RALL) Differential Input Impedance >2.5 M
Measuring Impedance Range 0.3~3.0Ω
Base line Impedance Range 0 ~4K
Bandwidth 0.1~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range Adult 0 ~ 120 rpm
Neo/PED0 ~ 150 rpm
Resolution 1 rpm
Accuracy ± 2 rpm
Apnea Alarm 10 ~ 40 S
NIBP
Method Oscillometric
Mode Manual, Auto, Continual
Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 (Min)
Measuring Period in STAT Mode 5 Min
Pulse Rate Range 40 ~ 240 bpm Alarm Type
SYS, DIA, MEAN Measuring and alarm range Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg
MEAN20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution 1mmHg
Accuracy Maximum Mean error ± 5mmHg
Maximum Standard deviation ± 8mmHg
Overpressure Protection Adult Mode 297± 3 mmHg
Pediatric Mode 240± 3 mmHg
Neonatal Mode 147± 3mmHg
SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 0 ~ 100 %
Accuracy 70% ~ 100% 2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec
Alarm Delay 10 Sec
Pulse Rate
Measuring and Alarm Range 20~300bpm
Resolution 1bpm
Accuracy ± 3bpm
TEMP
Channel 1
Resolution 0.1
Accuracy ± 0.2
Actualization interval about 1 Sec
Average Time Constant<10 Sec
Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature