Portable multi-channel temperature and humidity field testing system

Brand Name DAYAO
Place of Origin China
Model Number DY-XJY01B
Temperature Range -200-1600℃
Usage Industrial
Theory Temperature Controller
Accuracy 0.02%

Product Features

DY-XJY01B 

temperature and humidity field testing system

1.Product usage

It is used for detecting the temperature and humidity field of environmental test facility, such as electrically heated drying oven, biological incubator,resistance box, temperature humidity chamber,and heat treating furnace.

2.Functional characteristics

This device consists of automatic checkout test equipment ,signal acquisition unit,master chip,and system software. Both computer automated inspection and off -line inspection are available. And it has the function of data storage. Equipped with a 10-inch screen operating interface and the Bluetooth wireless, it is capable of automatic measurement.8000 data can be stored while automatically measuring. These data can be queried and uploaded to computer via USB or the Bluetooth wireless.

Weight: 6KG, Dimensions:500mm*360mm*160mm,Power supply: lithium battery (8.4V6AH) After charging full,it could be used for 24 hours.

This device will perform a calculation as to temperature deviation, volatility,and uniformity according to the relevant criteria. It also carries on uncertainty evaluation and do errors correcting from standard on a regular basis.Simultaneously, you can get the record ,certificate, and typing servicewhich will greatly improve work efficiency.

◆  Exquisite and nice appearance,Easy to carry, clear operational interface.

◆ This device consists of 15 temperature test channels and 4 humidity test channels,which can be increased in accordance with customers requirements. Uses imported humidity sensors to detect humidity.The reliability is very good.Thermocouple cold-end compensation system will compensate temperature automatically.

◆ A large collection of thermometric element , workableness thermocoupleSKNEJT thermal resistancept100Cu100Cu50Temperature Range-200~1600℃。

In the process of measuring,temperature value simultaneously shown by thermocouple and thermal resistance can be compared with its counterparts,namely MV or Ω .It is very convenient for users to make comparison.

Technical Index

 basic configuration

 measuring range

 precision

 Resolution

Automatic Data Processor

-200℃~1600℃

0.02%

0.01℃

  TC

Model K

(0~1100)℃

 superior to 1.5℃

0.01℃

Model T

(0~500)℃

 superior to 0.4℃

0.01℃

Model S

(300~1200)℃

 superior to 1.5℃

0.01℃

Pt1000

-50℃~300℃

 superior to 0.15℃

0.01℃

 RTD

Pt100

(-50~300)℃

 superior to 0.15℃

0.01℃

Cu50

(-50~150)℃

 superior to 0.5℃

0.01℃

 humidity testing

1%RH~100%RH

1%RH/ superior to 2%RH

0.1%RH

 4、 Products Show

 

       

 

5Detection Method And Referring To The Regulation

JJF1101-2003 calibration standards for temperature and humidity environmental testing equipment

JJF 1376-2012 calibration standards for chamber type electric resistance furnace

JJF 1171-2007 calibration standards for Temperature Circuit Measuring Instrument

JJF 1564-2016 calibration standards for Temperature and humidity standardizing box

Our company offer assistance in inspection service and issue a test report that complies with the relevant procedures.