H2S-A1 Hydrogen Sulfide Sensor (H2S Sensor)
H2S-A1 Hydrogen Sulfide Sensor (H2S Sensor)

H2S-A1 Hydrogen Sulfide Sensor (H2S Sensor)

USD $35 - $50 /Piece

Min.Order:1 Piece

Supply Ability:
3000 Piece / Pieces per Month
Port:
HK
Payment Terms:
T/T PayPal

Quick Details View All >

Brand Name:
Alphasense
Place of Origin:
United Kingdom
Model Number:
H2S-A1
Usage:
Gas Sensor
Theory:
Other, Electrochemical Gas Sensor
Output:
Analog Sensor

Week technology ltd

Business Type: Distributor/Wholesaler
Shenzhen Guangdong China
Main Products: Gas Sensor ,Temperature Sensor ,Humidity Sensor ,Pressure Sensor ,Ultrasonic Sensor ,UV Sensor ,Wind Speed sensor ,Liquid Level Sensor

Product Details

H2S-A1 Hydrogen Sulfide Sensor H2S Sensor Description
Alphasense Hydrogen Sulfide gas sensors operate using well-known fuel cell technology.

H2S-A1 Hydrogen Sulfide Sensor size:
20mm diameter, the industry standard size for portable gas detectors

With very high sensitivity and low zero, accurate and stable readings are possible when measuring current concentration ranges from ppb to 1% H2S.

H2S-A1 Hydrogen Sulfide Sensor H2S Sensor Schematic Diagram


H2S-A1 Hydrogen Sulfide Sensor H2S Sensor PERFORMANCE
Sensitivity
nA/ppm in 20ppm H2S
550 to 875
Response time
t90 (s) from zero to 20ppm H2S
<35
Zero current
ppm equivalent in zero air
<± 0.4
Resolution
RMS noise (ppm equivalent)
< 0.05
Range
ppm H2S limit of performance warranty
100
Linearity
ppm error at full scale, linear at zero and 20ppm H2S
0 to -4
Overgas limit
maximum ppm for stable response to gas pulse
500



H2S-A1 Hydrogen Sulfide Sensor H2S Sensor LIFETIME
Zero drift
ppm equivalent change/year in lab air
<0.1
Sensitivity drift
% change/year in lab air, monthly test
<3
Operating life
months until 80% original signal (24 month warranted)
> 24



H2S-A1 Hydrogen Sulfide Sensor  H2S Sensor ENVIRONMENTAL
Sensitivity @ -20°C
% (output @ -20°C/output @ 20°C) @ 20ppm
80 to 92
Sensitivity @ 50°C
% (output @ 50°C/output @ 20°C) @ 20ppm
100 to 110
Zero @ -20°C
ppm equivalent change from 20°C
<± 0.5
Zero @ 50°C
ppm equivalent change from 20°C<± 0.7



H2S-A1 Hydrogen Sulfide Sensor H2S Sensor CROSS  SENSITIVITY
NO2 sensitivity
% measured gas @ 10ppm NO2
<-20
Cl2    sensitivity
% measured gas @ 10ppm Cl2
< -25
NO    sensitivity
% measured gas @ 50ppm NO
<4
SO2  sensitivity
% measured gas @ 20ppm SO2
<10
CO    sensitivity
% measured gas @ 400ppm CO
<1.5
H2     sensitivity
% measured gas @ 400ppm H2
<0.2
C2H4 sensitivity
% measured gas @ 400ppm C2H4
<0.5
NH3   sensitivity
% measured gas @ 20ppm NH3
<0.1



H2S-A1 Hydrogen Sulfide Sensor H2S Sensor KEY SPECIFICATIONS
Temperature range
°C 
-30 to 50
Pressure range
kPa
80 to 120
Humidity range
% rh continuous 
15 to 90
Storage period
months @ 3 to 20°C (stored in sealed pot)
6
Load resistor
Ω (recommended)
10 to 47
Weight
g
<6



Matched Transmitter board for H2S-A1 H2S Sensor

4-20 mA Digital Transmitter Board
   • Factory calibration, custom-set for immediate use
   • Digital zero, sensitivity and temperature compensation
   • Digital and 4-20mA output
   • Biased or unbiased operation
   • Calibration and range change

   

Contact Supplier

Ms. Cherry sales Chat Now
Telephone
86-0755-83289036-843
Mobile
86-18086507787
Fax
86-0755-83289052
Skype
Isweek-Cherry Zhang
Address
16 Floor, Building #3, Zhongke Mansion, Gaoxin South Shenzhen,Guangdong
Whatsapp
18086507787

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Electronic Components & Supplies > Active Components > Sensors
Categories
Hi~