CL2-A1 Chlorine Sensor (CL2 Sensor)
USD $60 - $80 /Piece
Min.Order:1 Piece
CL2-A1 Chlorine Sensor (CL2 Sensor) Description
Alphasense Chlorine gas sensors operate using proven fuel cell technology.
CL2-A1 Chlorine Sensor size:
20mm diameter, the industry standard size for portable gas detectors
The Cl2 range provides OEMs with reliable sensors for use in a number of high volume applications. Strong signal levels combined with low zero current allows resolution to 20ppb and an operating range from +0 to 20ppm. The sensors are suitable for inclusion in fixed installation sensing heads and portable safety instruments.
These sensors have improved transient response to humidity changes compared to competitors' equivalent sensors.
The Cl2 sensors offer our electrolyte leak-free guarantee and reliable long-term detection performance. The leakproof housing is moulded with a colour coded top for ease of identification.
CL2-A1 Chlorine Sensor (CL2 Sensor) PERFORMANCE Sensitivity nA/ppm in 10ppm Cl2 -350 to -750 Response time t90 (s) from zero to 10ppm Cl2 (33Ω load resistor) <60 Zero current ppm equivalent in zero air ± 0.3 Resolution RMS noise (ppm equivalent, 33Ω load resistor) <0.02 Range ppm limit of performance warranty 20 Linearity ppm error at full scale, linear at zero and 5ppm Cl2 ± 1.5 Overgas limit maximum ppm for stable response to gas pulse 50 CL2-A1 Chlorine Sensor (CL2 Sensor) LIFETIME Zero drift ppm equivalent change/year in lab air, monthly test <0.05 Sensitivity drift % change/year in lab air, monthly test <10 Operating life months until 80% original signal (24 month warranted) > 24 CL2-A1 Chlorine Sensor (CL2 Sensor) ENVIRONMENTAL Sensitivity @ -20°C % (output @ -20°C/output @ 20°C) @ 10ppm Cl2 65 to 85 Sensitivity @ 50°C % (output @ 50°C/output @ 20°C) @ 10ppm Cl2 105 to 125 Zero @ -20°C ppm equivalent change from 20°C <± 0.2 Zero @ 50°C ppm equivalent change from 20°C <0 to -0.8 CL2-A1 Chlorine Sensor (CL2 Sensor) CROSS SENSITIVITY H2S sensitivity % measured gas @ 20ppm H2S <-300 NO2 sensitivity % measured gas @ 10ppm NO2 100 NO sensitivity % measured gas @ 50ppm NO <3 SO2 sensitivity % measured gas @ 20ppm SO2 <-8 CO sensitivity % measured gas @ 400ppm CO <0.1 H2 sensitivity % measured gas @ 400ppm H2 <0.1 C2H4 sensitivity % measured gas @ 400ppm C2H4 <0.1 CL2-A1 Chlorine Sensor (CL2 Sensor) KEY SPECIFICATIONS Temperature range °C -20 to 50 Pressure range kPa 80 to 120 Humidity range %rh continuous 15 to 90 Storage period months @ 3 to 20°C (stored in sealed pot) 6 Load resistor Ω (for optimum performance) 33 Weight g <6
Matched Transmitter board for CL2-A1 CL2 Sensor
4-20mA Digital Transmitter Board
• Factory calibration, custom-set for immediate use
• Digital zero, sensitivity and temperature compensation
• Digital and 4-20mA output
• Biased or unbiased operation
• Calibration and range change