asphalt mixture testing machine at high and low temperature
USD $4500 - $4500 /Set
Min.Order:1 Set
Jinan Heng Rui Jin Testing Machine Co., Ltd
WDW-LQ50/WDW-LQ100,50kN/100kN asphalt mixture testing machine at high and low temperature
1Introduction
1) tension, compression, bending testing and researching for asphalt and asphalt mixture at high and low temperature of -15℃~+60℃.
2) modified asphalt, elastomer modified asphalt, asphalt base, oil asphalt, minge asphalt, hydraulic oil asphalt, emulsified asphalt, hot mixed asphalt mixture, cool mixed asphalt mixture, modified asphalt mixture, flexible base, half rigidity base, rigidity base, cold mixed asphalt mixture, asphalt pavement, allowable tensile stress, tensile strength structural coefficient.
2 Important test:
1) T04713 - 2000 asphalt mixture single axial compression test(cylinder method: ¢100*100): deflection elastic modulous(20℃±0.5℃), stress of compression elastic modulous(15℃±0.5℃), compression strength, ball shape supporting, movable pressing head, loading:0.1p -- 0.7p, loading speed, variation coefficient.
2) T0714 -- 1993 asphalt mixture single axial compression test(prismoid 40*40*80): LVDT transducer, Max. breaking load, compression stress, strain and deformation, compression breaking modulous, asphalt mixture cracking property at low temperature(-10℃±0.5℃).
3) T0715 - 2011asphalt mixture bending test(15℃±0.5℃): (prismoid method: 40*40*80) bending test, asphalt mixture tensile property at low temperature(-10℃±0.5℃), LVDT transducer for displacement, beam type of suppportor/movable pressing head, deflection in the middle, Max. Breaking load, curve of deformation, bending modulous of elastic period.
4) T0716 - 2011 asphalt mixture ripping test (15℃±0.5℃): cylinder (¢100*40, ¢101.6*63.5mm, ¢150mm*40mm), crack strength test with asphalt mixture at low temperature(-10℃±0.5℃), vertical deformation, horizontal deformation 0.001mm, curve of load -- horizontal deformation, curve of load - vertical deformation, split tensile strength, tensile strain, breaking modulous, possion ratioμ, Max. Load, sepcific value of vertical deformation and horizontal deformation, total deformation at vertical direction.
3 Important specification
No. | Item | Specification |
1 | load capacity | 50kN / 100kN |
2 | Grade of testing machine | Grade of 0.5 / 1 |
3 | measure range of loading | Grade 0.4% - 100%F.S(0.5) /1% --100%F.S(grade 1) |
4 | relative error of indicating value | ±0.5%grade (0.5) /±1.0%(grade 1) |
5 | resolution of load | 1/±300000F.S(resolution is not changed in whole stroke) |
6 | range of deformation | 0.2% -- 100%F.S |
7 | adjusting range of speed for loading | 0.005% -- 5%F.S/S |
8 | control precision of speed of loading | When speed is <0.05%F.S,between ±1% of setting value. When speed is ≥0.05%FS,between ±0.5% of setting value. |
9 | adjusting range of speed for deformation | 0.005 -- 5%FS/s |
10 | control precision of speed for deformation | When speed is <0.05%FS/s,between ±1% of setting value When speed is ≥0.05%FS/s,between ±0.5% of setting value |
11 | resolution of deformation | 0.001mm |
12 | adjusting range of speed for displacement | 0.001 -- 600mm/min |
13 | , control precision of speed for displacement | Between ±0.2% of setting value |
14 | control range of constant loading,strain,displacement | 0.5% -- 100%F.S |
15 | control precision of constant loading, strain, displacement | When setting value is ≥10%F.S,betwee ±0.5% of setting value, When setting value is <10%F.S,between ±1 of setting value |
16 | moving stroke of transom | 1400mm / customized |
17 | effective testing span | 610mm / customized |
18 | Voltage(reliable landing) | Testing machine: 220V±10%, 50Hz environmental chamber:380V±10%, 50Hz |
19 | rate | 2.0kW+7.0kW |
20 | Size (LWH) | 1080*2100*2320mm |
21 | weight | 1800kg |
high precision environmental chamber for high and low temperature (moving type for long time test) | ||
22 | working temperature/ precision of temperature | -15°C -- +60°C/±0.5 °C |
23 | heating/refregiration /cooling mode | Resistance wire heating/(high grade of compensation), liquid nitrogen injection, air cycling |
24 | warming/cooling time and speed | RT+10°C-- 60°C/40min, RT+10°C-- -15°C/20min |
25 | regulator of temperature | Intellgent PID / displaying sensitivity 0.1°C/ temperature is not overshoot, fluctuation is small |
26 | size of woring room | 450*500x*400mm /customized |
27 | type of container | Made of stainless plate,front opening,with observing window,light in container. |