4OXV CiTiceL Oxygen (O2) Gas Sensor Technical Specifications
MEASUREMENT | |
Technology | Electrochemical |
Measurement Range | 1-25% vol. O2 |
Maximum Overload | 30% vol. O2 |
Output Signal | 0.10 ± 0.02 mA in Air |
Response Time (T90) | <15 Seconds |
Zero Current (Offset) | <0.6% vol. O2 |
Linearity | Can be considered linear in many cases. |
ELECTRICAL | |
Recommended Load Resistor | 100 Ω |
MECHANICAL | |
Casing Material | ABS |
Weight | <16 g |
Orientation Sensitivity | <0.2% vol. O2 equivalent |
ENVIRONMENTAL | |
Operating Temperature Range | -20°C to +50°C (up to 3 months continuous across RH range) |
Recommended Storage Temp | 0°C to 20°C |
Thermal Transient (Temp. plunge +22°C to -20°C) | <23.5% vol. O2 |
Operating Pressure Range | Atmospheric ± 20% |
Pressure Coefficient | <0.02% signal/mbar |
Pressure Transient (60 cm H2O step change) | <200% signal change |
Relative Humidity Range | (at 0°C to 20°C) |
Continuous | 5 to 95%RH non-condensing |
Short Term | 0 to 99%RH non-condensing |
LIFETIME | |
Long Term Output Drift | <2% signal/month Typically <5% over operating life |
Expected Operating Life | Minimum 24 months in air |
Storage Life | 6 months in original packaging |
Warranty Period | 24 months from date of despatch |
4OXV CiTiceL Oxygen (O2) Gas Sensor IMPORTANT NOTES
Obstruction of the vent hole will compromise performance, and must be avoided. Do not remove label.
Adequate free volume should be provided below the sensor to ensure full performance.
All tolerances ±0.15 mm unless otherwise stated.DO NOT solder to pins.
All performance data is based on conditions at 20ºC,50% RH and 1013 mbar
4OXV CiTiceL Oxygen (O2) Gas Sensor Typical Applications
General purpose, portable or fixed life safety.