60kHZ high frequency radiography X-ray machine PLD5800B
USD $1 - $1000000 /Set
Min.Order:1 Set
60kHZ high frequency radiography X-ray machine PLD5800B
Features:
§ The console applies LCD display photograph term
§ The console control the expose time by MCU internal clock, make the expose time accurate
§ Adopt microcomputer monitoring start-up and operation
§ Adopt mA,KV,s three parameters free matching work method
§ The low-voltage power supply was isolated in this machine
§ Hand-switch control module
§ External interlock device interface
§ This machine has temperature control switch to protect the x-ray tube
Basic Configuration:
1. 630mA control console one unit
2. High frequency high-voltage X-ray generator 60KHZ one set
3. PLD5800B radiography table one set
4. X-ray tube components one unit
5. Beam limiting device one set
6. High-voltage cable one unit
Optional Configuration
1. Chest stand (with JPI filter grids)
2. Bucky stand
Specification:
Item | content |
Output power | 50kW(max) 19kW(min) |
X-ray tube | Rotary anode, dual-focus, large focus: 1.2mm, small focus: 0.6mm, Tube thermal capacity: 900kJ |
Tube current | 32mA(min) ~630 mA(max) |
Tube voltage | 40kV-150kV with 41 steps |
Expose time | 1-6000ms |
Radiography grids | Grid density:103L/INCH;Grid ratio:10:1;Focusing distance:120cm; Stationary type:15"×18" |
Power supply | 380V±38V three phases, Five lines, 50Hz, Capacity: ≥40kVA, |
Structural performance:
Table size | 2000×760mm |
Table height | ≤700mm |
Table transverse movement | ≥220mm |
Table longitudinal movement | ≥900mm |
Grids cassette holder movement | ≥500mm |
X-ray tube assembly center rotation | ±90° |
Rotation of the pillar | 4×90° |
SID | 500-1280mm |
Pillar movement along the table | ≥1300mm |
Limited time of analog light beam opening | About 15s |
Film size | 5"×7"-14"×17" |
Chest stand up and down movement | 480~1460mm |