Specification
Channel: Single
Compatible syringe: 10ml 20ml 30ml 50/60ml
Infusion rate: 0.1-400ml/h, (50ml/60ml)
0.1-200ml/h, (20ml/30ml)
0.1-99.9ml/h, (10ml)
Purge rate/bolus rate/top infusion rate: 400ml/h; (50ml/60ml)
200ml/h; (20ml/30ml)
99.9ml/h; (10ml)
Volume limit: 0-999.9ml, Step by 0.1
Accuracy: +2%
Infusion Step by: 0.1ml/h
Visible & audible alarms: Syringe dislocation, Near empty, Pause overtime, Low battery, VTBI completion, Power cord falling, Malfunction, Incorrect syringe type, Electrical circuit jam
Occlusion pressure: High 20ml\\\\30ml\\\\50ml/60ml Syringe Space: 90kpa+30kpa
10ml Syringe Space: 100kpa+30kpa
Occlusion pressure: Low 20ml\\\\30ml\\\\50ml/60ml Syringe Space: 50kpa+30kpa
10ml Syringe Space: 70kpa+30kpa
KVO rate: 0.1ml/h+0.02ml/h
Power: AC: 100V-240V 50Hz/60Hz DC: 12V
Pause overtime alarm: 1min50s-2min
Appalled standards: ISO 13485 & CE Certified
Dimension: 290mm(L)*130mm(W)*130mm(H)
Weight: 2kg