KT-1 Particle Counter
Negotiable /Piece
Min.Order:1 Piece
Tianjin Tianda Tianfa Technology Co.,Ltd
It can be used for:
Hydraulic oils
Insulation- and turbine oils
Gear- and motor oils
Water based hydraulic fluids
Organic fluids
Features:
1. Size of the particle can be self-determined .
2. Using the injection pump as the motivity, see that the sampling is steady and veracity, along with the dopes degree of the sample does not impact on the speed of sampling.
3. Automated operating mode and manual operating mode can be selected ,the automated operating mode available for continually sampling and measuring 1 to 5 times.
4.The volume of the sample can be determined with the range of 1ml to 100ml,and the volume of the testing can be displayed on the real time .
5. Simple and convenient parameters setting and operation by using the few keys with a large touch screen display and user-friendly menu prompts. You can also chose the stander keyboard which is connected with the tester to run the entire test.
6. On switching on the power and starting the measuring, the tester automatically performs a self-checking and gives a status report displayed on the large touch screen.
7. Automated sampling pipe cleaning. A reverse flow through the sampling pipe is available for eliminating block up.
8. Easy calibration of the volume of the sampling liquid.
9. A standard printer port is available for connecting a printer to print out the measured results as well as actual measuring date and time.
10. The testing parameters can be automatically stored and reused for next test run
11. Good anti-interference feature. Reliable, precise and good repeatable measuring results.
12. The tester can test samples in-situ in sealed containers, beakers and bottles, With The probe rising and declining automatically
13. You can use the stander RS-232 interface to connect to the PC, with the help of the KT-1 software ,all sorts of function can be finished and the test result can be stored in the data-base of the pc.
14. The grades of the hydraulic pressure oil can be reported according to its stander .
Specifications:
1. Light Source: Semiconductor Laser
2. Measurement Range: 1μm-450μm
3. Sensitivity: 1μm(ISO 4402) or 4μm (C) (ISO 11171GB/T18854-2002)
4. Measuring Channel: 8 Random Channels
5. Sampling Volume Accuracy: Superior to ±1%
6. Flow Rate: 5mL/min~60mL/min
7. Pressure Range: 0~7bar
8. Samples Viscosity: ≤350cSt
9. Resolution: <10%(ISO11171GB/T18854-2002)
10. Max. Particle Concentration: 10000p/mL (5% coincidence error)
11. Operation Temperature: 0~50
12. Oil Temperature: 0~80
13. Data Output: Built-in Printer, RS232 Computer Interface
14.Power: 100~245V, 50HZ±1%, <80W