Child OLED Fingertip Portable Spo2 Pulse Oximeter For Hospital And Clinic
Negotiable /Kilogram
Min.Order:1 Kilogram
Child OLED Fingertip Portable Spo2 Pulse Oximeter For Hospital And Clinic
Product Configuaration:
Function | Configuration |
Display | OLED |
Spo2 parameter measurement | Y |
Pulse rate parameter measurement | Y |
PI parameter measurement | Sel |
Pulse waveform display | Y |
Bar graph display | Y |
Pulse beat show function | Y |
Four direction screen overturn function | Y |
Screen automatic overturn function | Sel |
Automatic power-off function | Y |
Abnormal parameter value alarm function with the measurement value sparkling | Y |
Abnormal parameter value alarm function with the sound | Sel |
Abnormal parameter value alarm function with the UI alarm string rolling | Sel |
Motion or light disturb alarm function with the UI alarm string rolling | Sel |
Arrhythmia indication function with the UI indication string rolling | Sel |
Low-power UI icon indication | Y |
Low-power UI character string alarm function | Sel |
Alarm setting function | Sel |
Pulse beat sound | Y |
Pulse sound On/Off function | Y |
Attention : \"Y\" expresses the function item is the standard configuration , and the \"Sel\" expresses the function item will be configured according to the users’ demand .
1.2 Users : Pediatric , adult.
1.3 Major applications and scope
Is used in measuring the pulse oxygen saturation and pulse rate through finger ; Is fit for family, hospital, oxygen bar, community healthcare, physical care in sports(It can be used before or after doing sports and it is not recommended to use the device during the process of having sport) and etc ; Not recommended to be used for monitoring the patients for a long time without any rest ; Do NOT use the device in an MR/CT environment, in an exprosive atmosphere, or on infant or neonatal patient . High temperature or high pressure steam disinfection for the oximeter is not permitted;Do not have the oximeter immerged in liquid ; Operation of this device below the minimum amplitude of 0.4% modulation may cause inaccurate results The device is designed to measure the percentage of arterial oxygen saturation of functional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following :Do not apply the pulse oximiter on the same arm as a blood pressure cuff, arterial catheter or infusion line(s)(IVs) .
Excessive light, such as sunlight or direct home lighting . Moisture in the device
Finger is outside recommended size range
Poor pulse quality
Venous pulsations
Anemia or low hemoglobin concentrations
Cardiogreen and other intravascular dyes
Carboxyhemoglobin
Methemoglobin
Dysfunctional hemoglobin
Artificial nails or fingernail polish .
Appearance
The above picture shows the appearance of EC-P60B , which is composed of main device, light emitter and the light sensor . The main device has such main components : button, OLED screen, battery component , and mechanical device which is divided into two parts .
Screen User Interface
The screen user interface has such information :
Spo2 parameter value and indication icon, 2-digit digital display ;
Pulse rate and indication icon , 3-digit digital display ;
Bar graph .
Pulse waveform .
Pulse beat icon .
Battery voltage indication icon .
Technical Specification
3.1 Power
Work voltage : DC 3V( 2 AAA Battery)
Work current : < 30mA
3.2 Screen Disply
Display screen : 0.96\" Dual-color OLED display(blue and yellow)
Screen Resolution : 128*64
Display Mode : four directions display modes can switch through shortly pressing the button under 2 seconds .
3.3 Spo2 Parameter performance
Range : 35~100%
resolution : 1%
Accuracy(>=70%):
Without disturb or low perfusion : +-2%
Motion/Light disturb or low perfusion : +-3%
Accuracy(<70%): below 70% unspecified .