1.HD LCD Display, high capacity words,friendly user interface,dynamically display working status.
2.Compatible with all brands syringes
3.Human voice alarm function: voice prompt and alarm will sound when infusion is almost done,infusion is finished,preset dosage is done,an injection blockage occurs,syringe device is improperly installed,a wrong speed is set,AC is unplugged,battery is close to zero.This further greatly reduces the workload of nurses(only available from our company)
4.Work Mode:ml/h and drops/min,can switch between these two models freely
5.Function to remove bubbles in the tube fast.
6.Function to handle fast speed and large volume(bolus)
7.The system automatically opens up KVO after infusion(reopens vein circulation)
8.Function to Preset Solution Volume:the user can preset solution volume.The machine will automatically stop and start KVO function once the preset solution volume is finished.
9.The machine automatically records the settings for last injection.It greatly reduces the workload for users.
10.Function to Auto-Define a Syringe:complete calibration technology for syringe brand definition that ensures the compatibility of any syringe brand and injection accuracy.
11.Three Power Source:AC220,50Hz,Internal Battery,12V car charge
Specifications: |
Speed Range
|
1ml/h~1200ml/h(increment of 1ml/h))
|
Infusion Accuracy
|
within±5% (accuracy after syringe is installed)
|
High Speed Infusion
|
100ml/h~1200ml/h(can be calibrated)
|
Volume Limit
|
1ml~999.9ml(increment of 0.1ml)
|
Infusion Volume Done
|
9999ml(increment of 0.1ml)
|
Alarm Pressure Value
|
High:800mmHg±200mmHg(106.7kPa±26.7kPa)
|
Medium:500mmHg±100mmHg(66.7kPa±13.3kPa)
|
Low:300mmHg±100mmHg(40.7kPa±13.3kPa)
|
Alarm Types
|
Preset volume is done,Infusion blockage,door is open,Bubbles in the infusion tube,Wrong infusion Speed,AC power unplugged,Battery is running out.
|
Power source
|
AC220V, 50Hz;Internal battery:power by DC11.1V rechargeable Lithium battery set,Capacity≥1600mAh;The battery can get this machine work at least 4hours at speed of 25ml/h (GB 9706.27-2005 Specified Medium Speed ) after charged for 10hours.
|
Bubble Detector
|
Ultra-sound type,sensitivity≥25uL
|
KVO Speed:
|
1ml/h~5 ml/h(increment of 1ml/h)
|
Power Consumption
|
30VA
|
Weight
|
1.8kg
|
Operation Environment
|
a) Temperature:+5℃~+40℃
|
b)Relative Humidity:20~90%
|
Transportation&Storage
|
a)Temperature:-30℃~+55℃
|
b)Humidity:≤95%
|
c)Atmospheric Pressure:50.10KPa~106.0KPa
|
Dimension
|
140mm(L)x 157mm(W)x 220mm(H)
|
Infusion Devices
|
Campatible with any brand
|
Forbidden
|
Not support blood Infusion
|
|