SN-41 Introductions:
This data logger can measure and store 4~20mAcurrent readings automatically. The professional applications can be providedby the PC software DGraph. Logger status, alarm indication and low batteryindication are via flashing red/green LED and yellow LED. The button on housingcan be used to start logging, or confirm and clear LED alarm when LED alarm ishold status.
User can plug datalogger straight into PC’s USB port without any cables. Before getting started,user need to install USB driver and DGraph software, provided on the CD whichis included in product packing box. All the software can run under Windows XP,Vista, 7 or 8(32 bit and 64 bit Versions). User can easily use DGraph softwareto setup sample rate, high/low alarm limit and start/stop method, and downloaddata to PC, etc.. And then user can view, analyse and print the data, or exportother file formats (txt, xsl, csv, bmp, jpg).
Especially, forsome professional applications, user can set access password, name eachchannel, enable/disable specified channel, input calibration data for eachchannel, disable LED indication for saving power, and select samplemode(Instant, Average, Maximum and Minimum), etc..
Features:
Portable, precise, compact and low cost
Easy to log and analyse data
Unique and uniform PC software DGraph can meet yourhigher requirements
Elaborate design makes it perfectly work in moreapplication field
Strict design ensure long term reliable work
Enhanced electrostatic protection design to preventunexpected damage
Each unit have single serial number identified byDGraph software to manage data simply
All data from different data logger and alldifferent measurement parameters can be graphed and analysed together
USB interface directly communicate with PC forsetup and data download
User can set access password for data logger
User can name each channel and enable/disablespecified channel freely
LED alarm enable/disable, LED alarm hold enable/disable and alarm limit settings for all channels via DGraphTM
User can input formula to customize compound line
User can create and customize new unit easily andfreely
Logging status indication via red/green LED andbattery status indication via yellow LED
Start logging method: Immediately, Button, Delayed
User can input two point calibration value for eachchannel when setup
Replaceable lithium battery and logged data are notlost if battery exhausted
Specifications:
DC Current Measurement range: 4 to 20mA
Accuracy(Typ.): ±1%±1count
Resolution: 0.05mA
LCD: With a LCD display
Memory Capacity: 64KBytes
Sample Rate: 1sec. to 12hr.
Start Method: Immediately, Start by button, andStart at date&time
Stop Method: Stop when memory full, overwrite whenmemory full, stop by button and stop until specified number of sample pointsfinished
Sample Mode: Instant (default), average, maximumand minimum, This will decide how to dispose data at idle time between a sampleinterval
Button: Software config to start logging by button,or clear and reset LED alarm when LED alarm is hold
LED: Red&green LED, Logger Statusindication(Include Delay to Start, Logging, Alarm, Finished)
(Softwarecan disable this two LED for saving power)
YellowLED, Low battery indication
Alarm :User programmable high or low alarm limit for each channel(If necessary); LEDalarm and hold
Communication: Direct USB connection; Need toinstall the driver provided on CD
PC Software: DGraph software and USB driverrequired (Included on CD)(Compatible with Windows XP, Windows Vista, Windows7&8(32 bit & 64bit Versions))
Power Supply: User replaceable 1/2AA 3.6V LithiumBattery 1200mAh
Battery Life: Approx. 3 Years @ 1 Minute SampleRate
Internal Input Impedance: 120Ω
Operating Range: 0 to 100%RH -35 to 80℃(-31 to 176℉)
Storage Range Recommended: 20 to 60%RH 10 to 50℃(50 to 122℉)
Case Material: ABS
Dimensions ;130mm×37mm×23mm(5.1"×1.5"×0.9")
Weight: Approx. 69 grams(2.43oz)
Standard Compliance: CE