GD-3536B Semi-auto Cleveland Open Cup Flash Point Tester (Digital Display)

GD-3536B Semi-auto Cleveland Open Cup Flash Point Tester (Digital Display)

Negotiable /Piece

Min.Order:1 Piece

Supply Ability:
0 Piece / Pieces per Month
Payment Terms:
T/T L/C D/P D/A
Delivery Detail:
7 days

Quick Details View All >

Brand Name:
Gold
Place of Origin:
China
Model Number:
GD-3536B

Chongqing Gold Machanical&Electrical Equipment Co.,Ltd.

Business Type: Trading Company
Main Products: oil purification ,oil filtering ,oil refinery ,oil regeneration ,oil recovery ,oil renewable ,oil disposal ,oil treatment ,transformer oil purifier ,lubricant oil recycling ,waste engine oil refinery ,turbine oil regeneration plant

Product Details

Purpose:

The instrument is designed and made as per the National Standard GB/T3536 "Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods), and T0651 "Asphalt Flash Point and Fire Point Test (Cleveland Open Cup Methods)" in the Industry Standard JTJ052 Specification and Test Methods of Bitumen and Bituminous Mixtures for Highway Engineering. It is used to determine flash point of petroleum products and bituminous materials, excepting fuel oil and materials having a flash point of lower than 79 centigrade.
 
Adopt microprocessor techniques and equipped with advanced system software, it can control temperature, apply flame, ignite, detect flash point, and print out test data automatically. It has a 6 digits display. 4 digits for temperature display and the other 2 digits for time display.

 

 

Technical parameters     

 (I) Operation condition
1. Ambient temperature: 20±5°C;
2. Relative humidity: < 80%;
3. Power supply: AC 220V±10%, 50±1Hz;
(II) Temperature measurement
1. Measurement range: ambient temperature~350°C;
2. Test accuracy: 0.2%;
3. Resolution:  0.1°C;
4. Repeatability:
(1) Flash point is lower than 150 °C: error is less than 4 °C;
(2) Flash point is higher than 150 °C: error is less than 6 °C;
(III) Basic parameters:
1. Speed rising speed: as per GB/T 3536-83;
2. Ignition mode: By a gas flame (4 mm);
3. Total power consumption: not more than 400 W;

Contact Supplier

Ren Pan Sales Chat Now
Telephone
86-023-62984892
Mobile
159-22636002
Fax
86-23-62940030
Skype
berurenpan
Address
JIntai Building Weihai,Shandong

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Measurement & Analysis Instruments > Testing Equipment
Categories
Hi~