3300 Series Rosemount Level Transmitter for liquid / solid , 4 - 20mA HART output
Negotiable /Piece
Min.Order:1 Piece
Beijing Zhongzhiguochuang Automation Equipment Co.,ltd
English Menu Portable ultrasonic Flow Meter with 3AAA Ni-H Battery
Rosemount 3100 Series Ultrasonic Level Transmitter 4-20mA with HART output
Handletype ultrasonic Flow Meter / flowmeter of High-accuracy measuring
lightweight precise 2051L Rosemount Level Transmitter of flange mounted
Rosemount 2051CD Differential Pressure measurement Transmitter , 0-0.3 to 0-10,000 psi
precision level / specific gravity measurement 3051L Rosemount Level Transmitter
3300 Series Rosemount Level Transmitter for liquids and some solids with 4 - 20mA HART communication output
Quick Detail:
· 3301 for liquids and some solids
· 3302 for liquids
· Time domain reflectometry (TDR) measurement principal
· SST, Monel, Hastelloy or PTFE-covered SST probe materials
· 4-20mA with superimposed HART communication output
· Up to 752°F process temperature
· Up to 5000 psi (345 bar) process pressure
· Intrinsic safety or Explosion/Flameproof Ex-Approvals
Description:
The smart, loop-powered Rosemount 3300 Series level and interface transmitter is based on guided wave radar technology. Reliable measurements on liquids and slurries are provided, even for severe conditions, due to advanced signal processing with digital sampling and a high signal to noise ratio.
Features of the Rosemount 3300 Series level and interface transmitter include easy installation and commissioning through two-wire technology, and pre-configured transmitter or user-friendly configuration with the included PC software package. Transmitter operation is virtually trouble-free as the Rosemount 3300 Series level and interface transmitter is unaffected by dust, vapor, interfering obstacles and turbulence. The Rosemount 3300 Series level and interface transmitter is also suitable for small or oddly shaped tanks. With direct level measurement there is no need for compensation for hanges in temperature, pressure, density, dielectric or conductivity. The robust modular design has a dual compartment housing allowing for removal without opening the tank.
The Rosemount 3300 Series level and interface transmitter can be used for level measurements on most liquids, semi-liquids, some solids and liquid / liquid interfaces. The Rosemount 3301 guided wave radar level transmitter is used for liquids and some solids. The Rosemount 3302 Multivariable ™ guided wave radar level and interface transmitter is used for liquids only.
Applications:
The guided wave Rosemount 3300 Series level and interface transmitter offer high reliability and performance. Measurements are virtually unaffected by temperature, pressure, vapor gas mixtures, density, turbulence, bubbling / boiling, varying dielectric media and viscosity. Since the waves are guided along the probe, this technology is excellent for small and narrow tanks / tank openings.
Specifications:
easurement Principle | Guided Wave Radar using TDR (Time Domain Reflectometry) Technology |
Maximum Measuring Range | 77 ft (23,5 m) from upper reference point |
Instrument Accuracy | ±0.2 inch (5 mm) for probes ≤16.4 ft (5 m) |
±0.1% of measured distance for rigid probes >16.4 ft (5 m) ±0.15% of measured distance for flexible probes >16.4 ft (5 m) |
|
Power Supply | Loop-powered (2-wire), 11-42 VDC (11-30 VDC in Intrinsically Safe (IS) applications, 16-42 VDC in XP applications) |
Output | 4-20mA with superimposed digital HART® communication Modbus |
Temperature/Pressure Rating | Refer to the Product Data Sheet |
Hazardous Location Approvals | ATEX: Flameproof and Intrinsic Safety |
CSA: Explosionproof and Intrinsic Safety | |
FM: Explosionproof and Intrinsic Safety | |
IECEx: Explosionproof and Instrinsic Safety | |
Display | Integral digital display (Option) |
Output Variables | Level, Distance, Interface Level, Interface Distance, Volume, Internal Temperature (Housing), Aout current and % of Range |
Enclosure | Polyurethane-covered die-cast Aluminum or stainless steel housing with dual compartments |
Ingress Protection | NEMA 4X, IP66 |
Competitive Advantage:
Accurate, direct level measurement virtually unaffected by process conditions as changes in temperature, pressure, density or conductivity and handles dust, vapor and interfering obstacles. Minimized maintenance with no moving parts, modular design and no re-calibration required. Fewer process penetrations and reduced installation costs with a MultiVariabl level and interface transmitter. Easy installation and commissioning through two-wire technology and user-friendly configuration. PC-based Radar Configuration Tools included for easy configuration, or optional preconfigured from factory. Easily integrated into existing HART and Modbus plant architecture, Wireless HART is optional by add one THUM adapter. High application flexibility with a wide range of process connectio