Product Details

High Accuracy Process Loop Calibrator with Quick-Click Interface And Rechargeable Battery , YHS-717

Brand Name HAODI
Place of Origin China
Model Number YHS-717

Product Features

Delivery Time:7-15days
Certification:ISO

High Accuracy Process Loop Calibrator with Quick-Click Interface And Rechargeable Battery , YHS-717

Quick Detail:

• Measures V dc to 28 V. 

• 0 - 20 mA or 4 - 20 mA default start up modes. 

• Push button 25% steps for fast,  easy linearity checks. 

• 1 µA resolution for mA source,  simulate and measure. 

• 0-100% “span check” for fast confirmation of zero and span. 

• HART ™ mode connects 250 Ω resistor in series with 

  24 V loop for compatibility with HART communicators.

• 24 volt internal loop supply,  so you can power and read

Description:

This loop calibrator is designed with output current,  Simulate output current,  output voltage function, it is used for detecting and debugging automatic control equipment in industry scene.

Features:

• Measures V dc to 28 V. 

• 0 - 20 mA or 4 - 20 mA default start up modes. 

• Push button 25% steps for fast,  easy linearity checks. 

• 1 µA resolution for mA source,  simulate and measure. 

• 0-100% “span check” for fast confirmation of zero and span. 

• HART ™ mode connects 250 Ω resistor in series with 

  24 V loop for compatibility with HART communicators.

• 24 volt internal loop supply,  so you can power and read 

  a transmitter at the same time without carrying a DMM. 

•Selectable slow ramp,  fast ramp,  and step ramp to

   provide smooth outputs for valve slewing and loop

• Innovative output adjustment dial with 1 µA and 100 resolution. 

• Single 9V rechargeable battery that is easily changed without braking calibration seal 

• Simultaneous mA and % readout for quick,  easy interpretation of readings. 

• mA accuracy of 0.025%,  superior to other loop calibrators in this price range. 

• Large display and simple,  quick-click interface for easy one-handed operation. 

functional tests.

Specifications

Current Measurement

Range:

0 to 24 mA

Resolution:

0.001 mA

Accuracy:

0.02% Rdg + 4 LSD

Current Sourcing

Range:

0 to 20 mA or 4 to 20 mA

Accuracy:

0.02% Rdg + 4 LSD

Drive Capability:

1200 Ω; 950 Ω at 20 mA in HART mode

Current Simulation

Range:

0 to 20 mA or 4 to 20 mA

Accuracy:

0.02% Rdg + 4 LSD

Ext. loop voltage requirement:

 12-30 V

Loop Power

 While measuring mA,  24 V

Voltage measurement

Range:

0 to 28 V DC

Resolution:

1 mV

Accuracy:

0.02% Rdg + 4 LSD