ISO standard electrolytic or infrared detection sensor method permeability tester
USD $1 - $1 /Piece
Min.Order:1 Piece
ASTM D3985 Best processing parameters of the oxygen gas permeability tester
ASTM standard Plastic films auto tensile tester, peeling force and strength tester
Pullout Test between Hypodermic Needle and Needle Cap, Sliding Resistance Test of Pistons
Medical packaging tensile and tear tester SYSTESTER supplier and manufactures
Automatic (SYSTESTER)water vapor permeability tester of plastic films
ASTM standard water vapor transmission rate tester of aluminum foils and laminated films
Brand: SYSTESTERInstruments co.,ltd
Standard test methods for water vapor and gas transmission rate orpermeability through polymeric membranes is based on the electrolytic orinfrared detection sensor method and available for the water vapor permeabilitytest of flexible films, sheeting and packages. It can test 3 specimens in oneoperation with independent results.
Features
2: Patent technology
3: Film and package test
4: 3 specimens testsimultaneously and independently
5:One-button test, testautomatically in whole process
6: Multiple test modes optional
7: Data curves display
9: Digital flow adjustment
10: 3 Chambers temperaturesensor
11: Environment temperature and humidity sensor inside
12: Chambers water bath technology
13: Unit based on embedded system,24bit Δ-Σ AD
14: Pressure, ppm water sensor and temperature overload protection
15: Traceable reference film calibration
16: Supports DSM system (DSM, lab data management system)
Technical indexes
Film: 0.001-52 g/m2·24h (Standard)
0.01-1100 g/m2·24h (Optional)
Package:
0.0001~0.5 g/pkg·d
Resolution:
0.00001 g/pkg·d
Specimen amount: 3specimens, independent
Humidity range: 0%RH,35%RH-90%RH,100%RH, standard condition 90%RH
Humidity accuracy: ±1%RH
Test area: 50 cm2
Specimen size:
Film: size: 150mm X 94mm
Thickness: less 5mm
Package:
Max size:D120mm X 400mm
Gas interface: 1/8 inch
Power supply: AC 220V 50Hz
Dimension: 730mm X 590 mm X 350mm
Net weight: 70kg
Standards
ISO 15106-3, GB/T 21529, DIN 53122-2, YBB 00092003
Configuration
Delivery: Main unit, W-TRANS software, Temperaturecontroller, tank regulator, PCIE card, communication cable, sample cutter,vacuum grease.