Product Details

PLX7000A High Frequency Mobile X-ray C-arm System (6KW)

Brand Name Perlong
Place of Origin China
Model Number PLX7000A

Product Features

1. Usage of c arm fluoroscopy

PLX7000A is wildly used in Department of Orthopedics,

Department of General Surgery, Orthopedics, Urology Surgery,

Spine Surgery, Abdominal Surgery, Department of Pain

Treatment, Cardiology, Gastroenterology, Gynecology,

Operating Room, etc.

 

2. Features of c arm fluoroscopy
1 High-power, micro-focus and high-frequency generator

optimize digital images.

2 Intelligent control of exposure makes low dose.

3 Many operating modes meet various clinical needs.

4 Muti-leaf and vertical light control reduces soft X-ray and

the skin dose effectively.

5 Imported well-known image intensifier and full-digital CCD

camera provide high-quality and high-resolution images.

6 Dual high-resolution LCD monitors assure image quality.

7 The powerful digital graphic workstation supports DICOM

3.0, the registration of worklist and manual registration.

8 Workstation has the function of high-capacity digital

storage. Fluoroscopy and digital spot film are stored in

lossless digital format. It has powerful processing

capabilities like edge enhancement, gamma correction,multiple

images, movie playback, window width/level, template experts,

burn, etc.

9 Four-D electric motion control, accurate positioning,

flexible and comfortable control, large frame design provides

lager diagnostic spaces and comfortable surgical environment.

New design and new ideas brings superior experience.

10 Two panels of human graphical LCD touch screen provide

intelligent and convenient operation. Dual motion control

system and double foot brake make this machine greatly meet

the demands of clinical operations.

 

3,Parameter of c arm fluoroscopy

 

High frequency inverter power supply
 Power output:6.0kW    Inverter Frequency:60 kHz
 
Automatic &Manual Continuous Fluoroscopy
 Tube voltage:40kV~125kV     adjust continuously
 
Tube current:0.3mA~4mA      adjust continuously
 
Automatic &Manual Enhanced Fluoroscopy
 Tube voltage:40kV~125kV     adjust continuously
 
Tube current:0.3mA~8mA      adjust continuously
 
DSI Digital Spotfilm Imaging
 optional 1-5
 
Photography tube voltage tube current
 40kV~125kV   120mA
 
X-ray tube special for high frequency
 Rotary anode focus 0.3/0.6 mm
 
Anode thermal capacity:212kJ