Product Details
The needle flame test is IEC60695-11-5, GB/T5169.5-2008,, and the GB4706.1-2005, standard use small flame light-off source program simulation test project
AUTO-ZY needle flame test instruments used in lighting, low voltage electric appliance, household appliances, machine tool appliances, motor, electric tools, electronic instrument, electrical instrument, electrical connectors, auxiliary items such as electrical and electronic equipment and parts, components of the research, production and quality inspection department, also suitable for insulation material, engineering plastic or other solid flammable materials industry.
Technical parameters:
1, needle burner is greater than 35 mm: high, stainless steel needle Φ 0.9 mm-Φ 0.5 mm.
2, burner Angle: 0 °, 20 °, 45 ° can be adjusted.
3:0 time was burning, 999999.9 s + /-0.1 s adjustable (generally choose for 30 s).
4, the burning time: 0-999999.9 s + /-0.1 s, automatic records, manual pause.
5, flame height: 12 mm ± 1 mm (with height measurement instruments).
6, and shows RanQiTi: more than 95% purity or butane propane.
7, temperature test range: 0 ~ 1000 ℃.
8, the flame temperature from 100 ℃ ± requirements: 2 ℃ went up to 700 ℃ ± 3 ℃ time in 23.5 seconds within ± 1 second.
9, temperature measurement thermocouple type: K Φ 0.5 mm precision plus or minus 0.05%.
10, test background: black background.
11, test process: test program automatic control, independent convulsions.
12, consult standard: IEC60695-11-5, GB/T5169.5-2008,, and the GB4706.1-2005.
13, studio: greater than 0.5 cubic volume, (but according to customers' special requirements 0.75 cubic or 1 cubic).
14, the equipment size: 1100 mm wide x 700 mm deep x 1300 mm high.
15, control: USES the monolithic microcomputer + touch screen control, the high accuracy thermometer independent type electrical control system, strong anti-jamming capability.
16, operation efficiency: test control processes to achieve a key type operation, fast and efficient test.
17, test standards according to actual test process integration: write test procedures
18, automatic: automatic ignition.