Product Details
1, burner: inner Φ 9.5 mm + /-0.3 mm Bunsen one.
2, test Angle: 0 °, 20 °, 45 ° manual switching.
3, flame height: 20 mm ± 2 mm to 180 mm ± 10 mm is adjustable.
4, ShiYan time: 0-999.9 s + /-0.1 s adjustable.
5, YuYan time: 0-999.9 s + /-0.1 s, automatic records, wire control pause.
6, burning time: 0-999.9 s + /-0.1 s, automatic records, wire control pause.
7, burning gas: 98% methane gas or 98% propane (generally can use liquefied petroleum gas (LPG) instead of).
8, flow pressure: with double flow and pressure gauge (gas and air)
9, temperature test range: 0 ~ 1000 ℃
10, the flame temperature requirements: from 100 ℃ ± 5 ℃ went up to 700 ℃ ± 3 ℃ time in 44 ± 2 seconds and 54 seconds ± 2 seconds.
11, temperature measurement thermocouple: K type Φ 0.5 mm precision plus or minus 0.05%.
12, test process: test procedures manual control, independent convulsions
13, consult standard: UL 94 2006, IEC 60695-11-3, IEC 60695-11-4, GB/T5169.15, 16, 17, 22-2008
14, studio volume: more than 0.75 cubic (can be customized to the other size)
15, the equipment size: 1200 mm wide x 700 mm deep x 1400 mm high
16, control: the single chip microcomputer + touch screen control, the high accuracy thermometer independent type electrical control system, strong anti-jamming capability.
17, operation efficiency: test control processes to achieve a key type operation, fast and efficient test.
18, testing standard integration: according to actual test process writing test procedures.
19, automatic: automatic ignition.