High Voltage Oil Breakdown Voltage Testing Equipment
USD $2,500 - $3,500 /Set
Min.Order:1 Set
Chongqing Nakin Electromechanical Co., Ltd.
High Voltage Oil Breakdown Voltage Testing Equipment
IIJ series fully automatic oil tester is designed as per the requirement of IEC156 Testing Method of Insulation Oil Medium Strength, which via proper programming, has got functions of auto boosting, step down, stirring, display and print out. After breakdown of oil medium, HV break off will operate within 10 seconds to enable an auto isolation of control system partly; this effectively safeguards the personnel security and equipment safety. The tester is suited for testing of the strength of insulating oil medium, as well as for testing of withstands voltage of transformer oil.
Features :
1. Fully automatic, control by microcomputer, utilization of new I/O terminal.
2. Numbers of times of testing and idling time of stirring are set via digital setting.
3. Test results displayed and printed out automatically.
4. Strong capability of anti-interference.
5. High accuracy of testing and high capacity of anti-jamming.
6. Easy to operate, small size, light weight which operated in the open air very conveniently and convenient for maintenance
7. Can store 99 groups oil testing data which is convenient to reading and print. The testing data can keep 100 years under the condition of power off.
8. System time can go on running under power off.(the system time just for reference can not be the standard of timing)
Technology and parameter data
1. Power supply: AC220V 10%,2. 50 HZ
3. Voltage output: 60 KV, 80KV,100KV
4. Capacity:1.6KVA, 2.0KVA
5. Pressure increase rate : About 2 KVA/S
6. Pressure testing speed: 3%
7. Break down sensitivity : <2KV
8. Wave form distortion :≤3%
9. Time of break down: ≤10ms
10. Operational environment: Temperature:0°C-40°C, Humidity: the most relative humidity85%
11. Storage environment : Temperature:-20°C-40°C, Humidity: the most relative humidity75%.
12. Operational height: <150m (can be specially designed if the height is over 1500m)
13.Test cell: Glass, volume 300 to 500 ml, with protective cover and stirrer.
14.Electrodes: Stainless steel spherical 36 mm diameter/hemispherical 25 mm radius, 2.5 mm gap as
per IEC-156.
15.Initial stand time of oil sample: 180 seconds
Intermediate stand time of oil sample: 60 seconds
String time of oil sample: 60 seconds
Number of consecutive test: 6
16.Weight for High voltage and control: Total about 55KG
NAKIN Company products list as following:
ZY-high effective vacuum transformer oil purifier, oil filtration
ZYD-Double stages Vacuum insulating/transformer oil purifier, oil treatment, oil purification
ZYM-Mobile/trailer type Vacuum transformer oil treatment, oil filtration machine
TY-Vacuum Turbine oil purifier, oil treatment Plant
TJ-Coalescence and separating light/fuel/gasoline oil filter, oil purifier machine
TYA-Vacuum Lubricating oil purifier/hydraulic oil purification
TYK- Phosphate ester fire-resistant oil purifier, oil filtration machine
JZS-Engine oil regeneration system, waste oil recycling
JZC-Waste Oil Distillation System
JYWL-Horizontal-closed type filtration device
TPF-Waste cooking oil recycling machine
BZ-Transformer oil regeneration device, oil purifier
PF-Plate pressure turbine/transformer/engine/diesel/hydraulic oil filter, oil purifier machine
JL-Portable turbine/lube/insulating oil purifier, oil filtering, oil filter
NKVW-Vacuum pump system
IIJ-II-BDV oil tester, insulating oil dielectric strength tester
NKEE-Oil moisture sensor, water content sensor