Medical mechanical ventilator systems/ICU ventilator
USD $1,500 - $2,000 /Unit
Min.Order:1 Unit
Wuhan Darppon Medical Technology Co., Ltd.
Medical mechanical ventilator systems/ICU ventilator PA500
Prospective service and support
The PA-500 ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. It is mainly intended for providing ventilation support to a critically ill patient during the life threatening phase and ensuring the going-through of the dangerous period by the patient and smooth treatment of primary diseases for recovery. Also it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation.
ALL MANIN UNIT HAS PASS THE CE The automatic conversion for standby power while AC power outage, keep working not less than 1 hour
Component Details
| Display: LED Display Following Parameters Operating mode and respiratory rate; ——Ventilator control frequency set by medical staff; ——Tidal volume of each breath in the patient; ——Occurrence of spontaneous breath in the patient and spontaneous breathing frequency; ——Expiratory phase and inspiratory phase and actual respiratory rate; |
Ventilator Construction: --Environmental Protection Material construction, hardly to be broken. -- Trolly: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions |
TECHNIQUES SPECIFICATIONS | |
Physical specifications | |
Display | LED Display |
Mode: | pneumatically powered and electrically controlled system |
Trolley: | Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions |
Technical Specifications: | |
Gas requirement: | Medical oxygen with a pressure ranging from 280~600kpa |
safety valve | <12.5 kPa |
Respiratory Rate | 6~60 times/ min |
Oxygen concentration output | <50% |
minute volume | ≥18L |
modes of ventilation | A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL |
PEEP: | 0.1 ~ 1.0 kPa |
Ptr: | -4 ~ 10 hPa |
IMV frequency: | 1~12 bpm |
I/E Ratio: | 1:1.5; 1:2.0; 1:2.5; 1:3.0 |
Tidal Volume | 0-1500ml |
Inspiratory Plateau: | 0~1s |
Maximum Safety Pressure | ≤ 6.0kPa |
Sigh: | one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point |
Pressure limit range: | 0 ~ 6.0 kPa |
Airway pressure alarm:
Audible and visual and with yellow and red color indicating | Lower: 0.9 kPa ~ 5.4 kPa; Upper: 0.5 kPa |
±0.2 kPa | |
Ventilation Volume alarm: | ≥ 18L/ MIN |
Power Supply Alarm | Ac/dc power supply are after failing to send out alarm immediately Alarm Time: keep >120s |
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator. |