R rotary evaporator series GREAT WALL
USD $1,853 - $8,321 /Set
Min.Order:1 Set
Zhengzhou Greatwall Scientific Industrial And Trade Co., Ltd.
R rotary evaporator series GREAT WALL
I. Introduction of rotary evaporator:
Large capacity, large diameter rotary evaporator flask, evaporation area, the vacuum water bath, while rotating edge heating, evaporation of the solution diffusion efficiency can be used for biological, pharmaceutical, chemical, food and other small scale areas, Chinese and production. Products with recycled water multi-use vacuum pumps, diaphragm pumps, low temperature cycle (vacuum) pumps, circulating cooler, constant temperature circulator, low temperature cooling liquid circulating pump and other devices supporting component systems.
II. Features of rotary evaporator:
1) Teflon (PTFE) and fluorine rubber dual rotating seal, patented technology to ensure high vacuum.
2) the introduction of foreign advanced frequency control system, temperature control system, accurate and reliable.
3) a vacuum switching valve, without affecting the system vacuum distillation of the situation and the solution can be continuously collected.
4) water bath electric lift and manual lift support, easy to use.
5) PTFE discharge valve, corrosion resistance, pollution-free.
III. Rotary evaporator-Vacuum-Low temperature system
20L,30L,50L middle and large type of rotary evaporator.
SHB-E big swept volume water circulating vacuum pump.
DLSB-30/30, 30/40, or DL-3000, DL-5000 of low temperature system.
Can provide the best solution for evaporate purified.
IV. Technical data:
Model | R-1005 | R-1010 | R-1020 | R-1050 | |||
Rotary Bottle | 5L,diameter of flange opening: Φ50mm | 10L,diameter of flange opening:Φ95mm | 20L, diameter of flange opening:Φ95mm | 50L,diameter of flange opening:Φ125mm | |||
Collecting bottle | 3L | 5L | 10L | 20L | |||
Speed-regulationj ofmain machine | Digitaldisplay,vayiable frequency,stepless Speed-regulation | ||||||
Rotary speed of mainMachine | 10to140r/min,digitaldisplay of rotary speed | 10to130r/min,digitaldisplay of rotary speed | 10to130r/min,digitaldisplay of rotary speed | 10to110r/min,digitaldisplay of rotary speed | |||
Rotaring motor (W) | phase AD induction motor,power:40W | phase AD induction motor,power:90W | phase AD induction motor,power:120W | phase AD induction motor,power:120W | |||
Condensing tube | Vertical type water refrigerating | Standing,main+auxiliary cooling and highly effcient3 circulating coldtrap | |||||
Heating boiler | stainless water bath | ||||||
Temperature control | Digital display,temperature control,0~99°C | ||||||
Achievable vacuum | 399.9Pa (below 3mmHg) | ||||||
Evaporating capability | Water≥2L/h Alcohol≥6.5L/h | Water≥3.2L/h Alcohol≥6.5L/h | Water≥5L/h Alcohol≥11L/h | Water≥9L/h Alcohol≥19L/h | |||
Elevating function | Electric elevation | Electric elevation | Electric elevation | Electric plus manual lifting | |||
Elevating stroke | 0~150mm | 0~160mm | 0~190mm | 0~230mm | |||
Rated power supply | 220V/50Hz | 380V/50Hz | |||||
Power of power supply | 2.2KW | 4.8KW | 6.3KW | 8.3KW | |||
Dimensions (mm) | 765L*400W*1070H | 920L*550W*1700H | 1250L*600W*2100H | 1320L*770W*2340H |
V. Recommended whole system (cooler + rotary evaporator + vacuum pump):
DLSB series cooler + R-1020 Rotary evaporator + SHB-B95 water circulating vacuum pump
VI. Common Problems and Solution of Rotary Evaporators:
Failure | Cause | Troubleshooting |
Power supply switched on but power indicator lamp not lit | Check the power line and find out whether it’s through or misconnected |
Connect the power line |
Failure of power switch | Stop using immediately and contact us | |
Failure of circuit board | ||
Power switch indicator lamp on but not in rotation | Bearing rusting | |
Failure of motor | ||
Failure of circuit board | ||
Nixie tube without display but still in rotation | Failure of nixie tube or transducer | |
Abnormal noise | Abrasion of seal ring | Replace the seal ring |
Abrasion of internal gear | Stop using immediately and contact us | |
Lack of oil in drive part | ||
Failure of motor | ||
Leakage of pressure and poor degree of vacuum | Abrasion of glass rotary shaft | Replace the glass rotary shaft |
Abrasion of seal ring | Replace the seal ring | |
Improper installation of seal ring (Opposite direction) | Remount the seal ring | |
Aging of the seal ring of pressure-relief air tap | Replace the air tap seal washer | |
Aging of vacuum hose | Replace the vacuum hose | |
Abnormal lifting | Failure of circuit board or motor |
Stop operating immediately and contact us |
VII. ISO9001:2008 Certificate:
VIII. Fairs: