Environmental Flow-type Salt Spray Test Chamber
USD $5,500 - $6,500 /Piece
Min.Order:10 Pieces
Guangdong Sanwood Environmental Chambers Co., Ltd.
Programmable salt spray chamber is applicable for various kinds material's surface treatment, like after coating , electroplate, inorganic and organic membrane dermails, negative pole with antirust oil treatment etc anti-corrosive treatment to test the corrosion-resistance of product..
A waterproof structure is adopted between the chamber cover and chamber body and thus there is no salt spray overflow. Adopts the tower spraying system and equipped with the salt solution filter system and non-crystal nozzle so as to realize even salt spray distribution and freely adjustable settlement. Design standard accrording:CNS:3627,3385,4159,7669,8886; JIS:D0201,H8502,H8610,K5400,Z2371; ISO: 3768,3769,3770; ASTM: 8117,B268.
Salt Spray Chamber is applicable for the salt spray corrosive test for the protection layer of components,parts, electronic and electrical parts and metal materials and industrial products. Salt Spray Chamber cover is made of transparent materials so that the operator can see the tested sample in the chamber and the spraying situation of the tested sample. Below are more details for programmable salt spray chambers !
Chapter 1: The main technical data for 270L SM-Y-90-D
01. | Performance: internal size: 900*600*500mm External size: 1470x1250x820mm |
02. | Laboratory temperature: Saline test(NSS ACSS)35℃±1℃/ Corrosion-proof test (CASS)50℃±1℃ |
03. | Pressure barrel temperature: Saline test(NSS ACSS)47℃±1℃/ Saline test(CASS)63℃±1℃ |
04. | The brine temperature: 35℃±1℃ 50℃±1℃ |
05. | Laboratory volume: 270L; Brine tank volume: 250L |
06. | Brine concentration: (CuCl2 2H2O)add 0.26 grams per liter of copper chloride to 5% sodium chloride solution or 5% of the concentration of sodium chloride solution |
07. | Compressed air pressure: 1.00±0.01kgf/cm2 |
08. | Spray volume: 1.0~2.0ml/80cm2/h (at least collect 16 hours and use the average value) |
09. | Laboratory relative humidity : more than 85% |
10. | PH Value: 6.5~7.2 3.0~3.2; Spray way: Programable spray |
11. | Power: AC220V 1Φ 20A |
Chapter 2: Key components for 270L SM-Y-90-D
01. | Solenoid valve, Brand: SMC 4 pcs |
02. | AC contactor, Brand: LG Korea with 1 pc |
03. | Temperature Sensors, Brand: Taiwan Songqi PT100 with 2 pcs |
04. | Intermediate relay, Brand: Japan Omron with 3 pc |
05. | Time Relay, Brand: Japan Omron with 1 pc |
06. | Water shortage protection, Brand: Taiwan with 2 pcs |
07. | PVC plate, Brand: Taiwan imported with 2 pcs |
08. | Solid State Relay, Brand: Japan Omron with 2 sets |
09. | Heater, Brand: China Feiyang with 2 sets |
10. | Buzzer, Brand: Taiwan Zhifeng |
11. | Circuit Breaker, Brand: Taiwan Shilin with 5 pcs |
12 | Temperature Controller, Brand: Japan Dahua with 2pcs |
13 | Saturated barrel, Brand: SUS #304 |
14 | Switch, Brand: Taiwan Tiande 5 pcs |
Chapter 3: Material Issue for SM-Y-90-D
01. | Laboratory is made of PVC polyvinyl chloride sheet (Germany), the thickness is 8mm, the temperature resist 80 ℃ |
02. | Carton with PVC polyvinyl chloride plates (Germany), the thickness is 8mm, the temperature resist 80 ℃ |
03. | Laboratory closures use impact-resistant acrylic plate (Germany), thickness is 8mm. |
04. | Reagent supplemented with hidden water level perspective window, easy to clean,hard to be broken. |
05. | Saturated air barrels adopt SUS # 304 stainless steel with high pressure barrel insulation. |
06. | Laboratory baskets use planar dividing frame, the angle can be adjusted, surrounded by the fog and mist so place most test piece. |
Chapter 4: CNS.JIS.ASTM specification, constant temperature control can be set
01. | Salt spray test: NSS, ACSS (1) Laboratory: 35 ℃ ± 1 ℃ (2) saturated air barrel: 47 ℃ ± 1 ℃ |
02. | Corrosion test: NSS, CASS (1) Laboratory: 50 ℃ ± 1 ℃ (2) saturated air barrel: 63 ℃ ± 1 ℃ |
03. | Standard according to IEC68-2-11, ISO7253,GB10587, GB/T2423.17, GB/T5170.8, GJB150.11-86. It suitable for all kinds of electrical materials and parts into the salt spray corrosion testing. The disposable molded fiberglass shell, transparent cover. |
04. | Temperature fluctuation ≤ ± 0.5 °C, temperature uniformity ≤ ± 2 °C. Salt settlement: 1 ~ 2ml/80cm2 hours, special nozzles, no crystallization phenomena. |
05. | High-precision digital meters, PID fuzzy control. Pt100 test sensors, dual pressure protection and over-temperature protection. Air-spray, spray and adjustable time interval can be set. |
Chapter 5. Air Supply System/ Spray Ways/Methods
01. | The air supply system: air pressure is 1kg/cm2 and adjusted by two-stage; The first stage just need rough adjustment, and the second stage need precision adjustment. |
02. | Adopting Bonute principle to absorb brine and then atomized with uniform atomization and no blocking crystallization, to ensure the continuous testing standards. |
03. | Nozzles: Adopting special glass (PMMA) nozzle, use time up to 10000H,spray volume and spray angle can be adjustable. |
04. | Spray volume: spray tower set in the middle of laboratory so that keep uniformed fog, you can debug 1-2ml 80cm2/hr (taking the average amount in 16 hours) |
05. | Continuous spray, spray cycle time setting function is adjustable. |
Chapter 6. Features of Salt Spray Chambers
01. | The spray nozzle the heart of tester adopt Bo Nute Principle which change saline water into atomization but no crystallization produced. |
02. | The controller is digital type microcomputer which delivers precision temperature |
03. | Parameters show on the display screen such as temperature, humidity, pressure, spray capacity and time. |
04. | Test specifications are accept worldwide. |
Salt Spray Test Chamber is applicable for various kinds material's surface treatment, such as after coating, electroplate, inorganic and organic membrane dermails, negative pole with antirust oil treatment etc anti-corrosive treatment to test the corrosion-resistance of the product . Welcome your enquiry ! Thank you.
Chapter 7. Heating System for Chambers
01. Direct heating with rapid heating speed so reduce the standby time, when the temperature reaches the preset value,ON / OFF switch can be change. So the temperature is accurate and less power consumption. (Heating tube is titanium alloy)
Chapter 8. Control System:
01. | The liquid expansion in heating tank overtemperature protection controller 0-120 ℃ |
02. | The liquid expansion in saturated airbarrels over-temperature protection controller 0-120 ℃ |
03. | Digital time display controller 0.1s-99990hr |
04. | Timer:0-9990hr |
05. | Rocker Illuminated Push Button Switch, Push Button can work continuously 20000 times |
Chapter 9. Water Supply System/ Safety Protection Device
01. | Adopt automatic water replenishment system, water complement automatically when the water level is too low. |
02. | When over-temperature, automatically cut off power supply unit. |
03. | With two safety warning light device. |
04. | When water level is too low, automatically cut off the power supply unit. |