Specifications
Two channel with two governer,
Occlusion Pressure Selected,
Anti bolus,
Syringe Brands Selected
AC and DC indicate
CHARACTERISTIC
- Automatically detection of syringe types and settings.
- Broad range in delivery rate settings.
- 8 Hours internal battery life in high capacity.
- Three levels of alarm volume: high, medium and low.
- Three levels of occlusion pressure: high, medium and low.
- KVO function: After infusion is complete, the pump automatically switches to the smallest flow to prevent creating the thrombus.
Compatible Syringe: Any standard syringe with 10ml,20ml(30ml) and 50ml(60ml)
Accuracy: ±2% ( Any manufacturers of syringes, right after calibration )
Delivery Rate Range:
50ml(60ml): 0.1ml/h~1800.0ml/h (0.1ml/h step)
20ml(30ml): 0.1ml/h~800.0ml/h (0.1ml/h step)
10ml: 0.1ml/h~400.0ml/h (0.1ml/h step)
Time Mode Range: Volume: 0.1~999.9ml (0.1ml step)
TIME: 1~1999min (1min step)
Body Weight Mode Range:
Volume: 0.1~999.9ml(0.1ml step)
DRUG MASS: 0.1~999.9mg(0.1mg step)
DOSE: 0.01~99.99ug/kg/min(0.01ug/kg/min step)
WEIGHT: 0.1~300kg(0.1kg the least step)
Volume Limit Range: 0~1999.9ml (0.1 ml step)
Total Volume Range: 0.1~1999.9ml (0.1 ml step)
Rapid injection: 50ml(60ml): 1800.0ml/h, 20ml(30ml): 800.0ml/h, 10ml: 400.0ml/h
KVO Rate: 1ml/h
Alarms: Near empty, Syringe empty, Volume complete(a volume limit has been set ), Occlusion, Syringe loose, Operation error, Low battery and other device failures.
Power Supply: AC power supply: AC 100 to 240V±10%,50/60Hz;
Internal Lithium Battery:11.1V, 2000mAh. Approximately 8 hours of continuous operation on a new and fully charged battery running at 5ml/h.
Power Consumption: ≤25VA
Safety level:ClassI, Type CF
Dimensions: 280mm(L)×160mm(W)×130mm(H)
Net Weight:Approximately. 3.0kg
Operating Conditions: Ambient temperature: 5 to 40°C Relative humidity: 20 to 90%
Storage Condition: Ambient temperature: -10 to 55°C Relative humidity: 10 to 90% Accessories: AC power line, Operational manual