Handheld Medical Multi-parameter Patient Monitor
Negotiable /Piece
Min.Order:10 Pieces
Powerful(All functions are as same as other JP series monitors)
6.4" Colour TFT LCD(640*480)
Inside Li-Battery and low power waste(Less than 4w)
Can work for 4 hours with battery
BD6000 series patient monitor is competent for the monitoring of the Echocardiograph, Heart Rate,Temperature,Blood Pressure,Respiration and SpO2.
Feature
1.Portable monitor for 24 hours measurement.
2.Analysis for arrhythmia and ST segment. Store 96 hours daily data and abnormal data.
3.E-Document instruction for operator.
Technical Specification
1.Resist electric shock: Class I Class CF
2.Apply to: Adult, Child, Baby and Resolution: 640*480
3.Display Waveform: 4 Leads Display waveforms of ECG, Respiration, NIBP SpO2
4.Colours of Waveform: 15 Kinds
5.Freeze Waveform: Enable
6.Monitor Functions: Heart Rate, VPC, Three Level ST Segment,NIBP, SpO2, Pulse Rate, Respiration Rate and Temperature.
ECG
1.Input: 5 Electrode Leads RA, LA, RL, LL, C (V1~V6)
2.Electrode Lead: Standard I, II, III; Body Electrode Lead aVR, aVL, aVF; Bosom Electrode Lead V1~V6
3.Signal Amplifying: 0.5, 0.75, 1.0, 1.5, 2.0
4.Range of ECG Measurement: 30 times/min~250 times/min
5.Precision of ECG: ±5%
6.Scan Speed: 12.5mm/s, 25mm/s, 50mm/s
7.ECG Review: Enable
8.Range of Alarm: 30 times/min~250times/min
9.Mode of Alarm: Sound and Screen Remind Alarm
10.Work with electrosurgical unit: Enable
11.Sustain 4000VAC/50Hz electric shock
NIBP
1.Measure Method: Advanced One-step Method with Automatic / Manual Measurement
2.Measure Unit: mmHg and kPa
3.Measure parameter: SYST(systolic pressure), DIAS(diastolic pressure), MEAN(mean arterial pressure)
4.Measure Range: 10mmHg~280mmHg; 1.3kPa~37kPa
5.Precision of NIBP: ±5mmHg;±0.7kPa
6.Protection of Over Pressure: 330mmHg; 44.0kPa
7.Alarm: SYST, DIAS, MEAN
SPO2
1.Range of SpO2 Measurement: 40%~99% with the distinction of 1%
2.Precision of SpO2: During 85%~99%:±2% ; During 60%~85%:±4%
3.Alarm Range: 0~99%
4.Measure Pulse Rate Range: 30 Times/min~200 Times/min
5.Precision of Pulse Measurement: ±5%
Temperature
1.Measure Range: 32°C~43°C with the distinction of 1°C
2.Precision of Temperature Measurement: ±0.2°C
Respiration
1.Measure Range: 4 Times/min~120 Times/min
2.Precision of Respiration Measurement: ±2 Times/min
3.Waveform Display: Display 4 Waveform in screen.
4.Trend of data: Store 96 hours chart of data of Heart Rate, NIBP, SpO2, Temperature, Respiration, VPC and ST Segment.
5.Store ECG Waveform: Store 64 waveforms of ECG by every 10 seconds. Can freeze the waveform.
Alarm
1.Message for the abnormal system alarm of lost lead or too much noise.
2.Message for the abnormal data alarm of Heart Rate(Tachycardia and Bradycardia), NIBP, Temperature, SpO2, ST Segment(ST Elevation and ST Depression), VPC, Bigeminal rhythm, Tirgeminal rhythm, RonT, VTAC, Escape Beat and Sinus arrest.
3.Ways of Alarm: Sound and message in the screen.
4.Alarm Level: 3 levels