CE approved Infusion pump with bolus function
Negotiable /Piece
Min.Order:5 Pieces
JAS-1200(JSB-1200)
FEATURES
*Compatible with IV set of any standard
*High accuracy and eliminates false alarm with unique design of drop sensor
* Three work modes: Drop counting control, volume control and time control mode
* 8 hours internal battery life in high capacity
* Special light for night operation
* Three levels of occlusion pressure
* Three levels of alarm volume
* Waterproof and anti-sticking design of the drop detector
* Bolus function
TECHNICAL SPECIFICATIONS
*Control way:Drop counting and adjusting the delivery rate automatically
*Pump Type:Peristaltic finger
*Applicable IV Sets: Compatible with IV sets of any standard (including lucifugal IV sets)
*Accuracy:≤25drop/min:±2%, >25drop/min:±3% (With the drop counting control mode)
±5% (volume control model with correct calibration)
*Delivery Rate Range:
1~400drop/min(1drop/min step)
1.0ml/h~1200ml/h (1.0ml/h~99.9ml/h, 0.1ml/h step; >99.9ml/h, 1ml/h step)
*Delivery Time Range:1min~1999min (1 min step)
*Volume Limit Range:0~9999ml (1 ml step)
*Total Volume Range:1~9999ml (1 ml step)
*Purge Rate:600ml/h
*Bolus:Bolus rate: 10-1200ml/hr, Bolus volume: Maximum of 10ml
*KVO Rate:When the delivery rate≤10ml/h, KVO is 1ml/h, otherwise KVO is 3ml/h.
*Alarms:Air-in-line, Infusion complete when limited volume has been set, Empty container, Occlusion, Door is open, waiting over time, No drop been detected, Operational errors, Low battery, other device failures.
*Power Supply:AC power supply: AC 100 to 240V±10%,50/60Hz;
Internal LI Battery: 11.1V, 2000mAh. 8 Hours of continuous operation on a new fully charged battery with the rate of 20 drops/min.
*Power Consumption:≤25VA
*Safety level:ClassI, Type BF
*Dimensions:130mm(L)× 200mm(W) × 230mm(H)
*Net Weight:Approximately. 2.5kg